| Literature DB >> 24335806 |
David Dornbos1, Promod Pillai1, Eric Sauvageau1.
Abstract
Small ruptured aneurysms present a unique problem to endovascular therapy. We report a case in which a patient presented with subarachnoid hemorrhage and a very small ruptured ophthalmic artery aneurysm, for which endovascular therapy was preferred secondary to severe cardiac comorbidities. Due to the aneurysm size, a small 1.5 mm coil was needed, but presented a significant risk of migration. Conventional stent assisted coiling was considered suboptimal as the small coil could have easily migrated through the strut. We present a novel technique of flow diverter assisted coil embolization in which a coil was placed within the aneurysm and a pipeline embolization device was then partially deployed, jailing the microcatheter and coil mass. Once in place, the coil was detached, securing the aneurysm, and preventing coil migration. Through the use of a flow diverter, some degree of aneurysm protection would still be expected in the event of coil migration toward the ophthalmic artery origin. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/Entities:
Keywords: Aneurysm; Angiography; Coil; Flow Diverter; Subarachnoid
Mesh:
Year: 2013 PMID: 24335806 DOI: 10.1136/neurintsurg-2013-010876.rep
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836