Literature DB >> 24335551

Limits of agreement between measures obtained from standard laboratory and the point-of-care device Hemochron Signature Elite(R) during acute haemorrhage.

T Gauss1, S Hamada, I Jurcisin, S Dahmani, L Boudaoud, J Mantz, C Paugam-Burtz.   

Abstract

BACKGROUND: Rapid diagnosis of coagulopathy in the bleeding patient using point-of-care (POC) devices would be ideal. The Hemochron Signature Elite(®) (HC(®)) is a POC device that determines international normalized ratio (INR) and activated partial thromboplastin time (aPTT). The aim of the study was to evaluate the agreement for INR and aPTT between the HC(®) and standard laboratory values in acute haemorrhage.
METHODS: This was a single-centre observational prospective study including patients with acute haemorrhage. Laboratory INR and aPTT were compared with simultaneous measurements performed with the HC(®). The diagnostic performance of HC(®) was determined; bias and limits of agreement were calculated according to the method of Bland and Altman.
RESULTS: Seventy-two pairs of measurements from 39 patients were analysed. The bias between the INR-HC(®) and aPTT-HC(®) measurements and the central laboratory were 0.02 and -1.13, respectively. The Spearman's correlation coefficients for the INR-HC(®)/INR-lab and the aPTT-HC(®)/aPTT-lab were 0.68 and -0.29, respectively. Twenty-seven per cent of INR-HC(®) values and 89% of the aPTT-HC(®) values exceeded the predefined limits of agreement. The INR-HC(®) measurement identified patients with a central laboratory INR >1.5 with a sensitivity, specificity, and positive and negative predictive values of 83%, 70%, 76%, and 77%, respectively.
CONCLUSIONS: The results showed a lack of agreement between the INR-HC(®) and the aPTT-HC(®) measurements and the standard laboratory in the context of acute haemorrhage. The INR-HC(®) showed moderate performance as a decision-making tool to detect coagulopathy in the context of acute haemorrhage.

Entities:  

Keywords:  blood coagulation disorders; haemorrhage; injuries; point-of-care systems; surgical blood loss

Mesh:

Year:  2013        PMID: 24335551     DOI: 10.1093/bja/aet384

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  9 in total

1.  All the bang without the bucks: Defining essential point-of-care testing for traumatic coagulopathy.

Authors:  Michael D Goodman; Amy T Makley; Dennis J Hanseman; Timothy A Pritts; Bryce R H Robinson
Journal:  J Trauma Acute Care Surg       Date:  2015-07       Impact factor: 3.313

2.  The European guideline on management of major bleeding and coagulopathy following trauma: fifth edition.

Authors:  Donat R Spahn; Bertil Bouillon; Vladimir Cerny; Jacques Duranteau; Daniela Filipescu; Beverley J Hunt; Radko Komadina; Marc Maegele; Giuseppe Nardi; Louis Riddez; Charles-Marc Samama; Jean-Louis Vincent; Rolf Rossaint
Journal:  Crit Care       Date:  2019-03-27       Impact factor: 9.097

3.  Activated clotting time in inpatient diagnostic and interventional settings.

Authors:  Heidi Dalton; Michael Martin; Pamela Garcia-Filion; David Shavelle; Pei-Hsiu Huang; Justin Clark; Sean Beinart; Andrew Israel; Nichole Korpi-Steiner
Journal:  J Thromb Thrombolysis       Date:  2022-06-22       Impact factor: 2.300

4.  The European guideline on management of major bleeding and coagulopathy following trauma: fourth edition.

Authors:  Rolf Rossaint; Bertil Bouillon; Vladimir Cerny; Timothy J Coats; Jacques Duranteau; Enrique Fernández-Mondéjar; Daniela Filipescu; Beverley J Hunt; Radko Komadina; Giuseppe Nardi; Edmund A M Neugebauer; Yves Ozier; Louis Riddez; Arthur Schultz; Jean-Louis Vincent; Donat R Spahn
Journal:  Crit Care       Date:  2016-04-12       Impact factor: 9.097

5.  Thrombolysis in stroke patients: Comparability of point-of-care versus central laboratory international normalized ratio.

Authors:  Ramona C Dolscheid-Pommerich; Sarah Dolscheid; Lars Eichhorn; Birgit Stoffel-Wagner; Ingo Graeff
Journal:  PLoS One       Date:  2018-01-10       Impact factor: 3.240

6.  Performance of point-of-care international normalized ratio measurement to diagnose trauma-induced coagulopathy.

Authors:  Thomas Mistral; Yvonnick Boué; Jean-Luc Bosson; Pauline Manhes; Jules Greze; Julien Brun; Pierre Albaladejo; Jean-François Payen; Pierre Bouzat
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2017-06-21       Impact factor: 2.953

Review 7.  Korean clinical practice guideline for perioperative red blood cell transfusion from Korean Society of Anesthesiologists.

Authors:  Bon-Nyeo Koo; Min A Kwon; Sang-Hyun Kim; Jong Yeop Kim; Young-Jin Moon; Sun Young Park; Eun-Ho Lee; Min Suk Chae; Sung Uk Choi; Jeong-Hyun Choi; Jin-Young Hwang
Journal:  Korean J Anesthesiol       Date:  2018-12-05

8.  Tools to predict acute traumatic coagulopathy in the pre-hospital setting: a review of the literature.

Authors:  Simon Robinson; Jordan Kirton
Journal:  Br Paramed J       Date:  2020-12-01

9.  An Agreement Study Between Point-of-Care and Laboratory Activated Partial Thromboplastin Time for Anticoagulation Monitoring During Extracorporeal Membrane Oxygenation.

Authors:  Yuan Teng; Shujie Yan; Gang Liu; Song Lou; Yang Zhang; Bingyang Ji
Journal:  Front Med (Lausanne)       Date:  2022-06-29