Literature DB >> 24334986

A new method for determining the absorbed dose in a radiation field by using a thiamine hydrochloride aqueous solution.

Ying Li1, Changyin Lv, Ying Zhao, Qiang Sun, Yunjing Li.   

Abstract

This research investigated the ionizing radiation effect on thiamine hydrochloride (TH) and its usability as an irradiation dosimeter. The fundamental principle is to determine the concentration variations of TH solutions with high-performance liquid chromatography (HPLC) after exposing to γ-rays. The decreasing peak area of TH in the HPLC chromatogram forms a linear relationship with the rising radiation dose. We investigated the characteristics and suitable application range of the TH as a new radiation dosimeter. The influence factors and mechanism of the reaction induced by radiation were also discussed. According to the correlation between the concentration and the radiation dose, 0.3 g/L of a TH solution is suitable for the 0.1-10 kGy dose range, and 2 g/L is appropriate for 0.1-20 kGy. The easy availability and the simple, but stable, chemical structure of thiamine makes it a potential candidate for radiation dose research. Also, the preparation proceeding for sampling is easy, and the result can be automatic monitored by liquid chromatography.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24334986     DOI: 10.2116/analsci.29.1189

Source DB:  PubMed          Journal:  Anal Sci        ISSN: 0910-6340            Impact factor:   2.081


  1 in total

1.  A New Fluorometric Dosimetry for Low-medium Gamma Radiation Doses.

Authors:  Ece Ergun
Journal:  J Fluoresc       Date:  2021-04-02       Impact factor: 2.217

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.