Effect of Acuros XB algorithm on monitor units for stereotactic body radiotherapy planning of lung cancer.

Rao F Khan1, Eduardo Villarreal-Barajas2, Harold Lau2, Hong-Wei Liu2.   

Abstract

Stereotactic body radiotherapy (SBRT) is a curative regimen that uses hypofractionated radiation-absorbed dose to achieve a high degree of local control in early stage non-small cell lung cancer (NSCLC). In the presence of heterogeneities, the dose calculation for the lungs becomes challenging. We have evaluated the dosimetric effect of the recently introduced advanced dose-calculation algorithm, Acuros XB (AXB), for SBRT of NSCLC. A total of 97 patients with early-stage lung cancer who underwent SBRT at our cancer center during last 4 years were included. Initial clinical plans were created in Aria Eclipse version 8.9 or prior, using 6 to 10 fields with 6-MV beams, and dose was calculated using the anisotropic analytic algorithm (AAA) as implemented in Eclipse treatment planning system. The clinical plans were recalculated in Aria Eclipse 11.0.21 using both AAA and AXB algorithms. Both sets of plans were normalized to the same prescription point at the center of mass of the target. A secondary monitor unit (MU) calculation was performed using commercial program RadCalc for all of the fields. For the planning target volumes ranging from 19 to 375cm(3), a comparison of MUs was performed for both set of algorithms on field and plan basis. In total, variation of MUs for 677 treatment fields was investigated in terms of equivalent depth and the equivalent square of the field. Overall, MUs required by AXB to deliver the prescribed dose are on an average 2% higher than AAA. Using a 2-tailed paired t-test, the MUs from the 2 algorithms were found to be significantly different (p < 0.001). The secondary independent MU calculator RadCalc underestimates the required MUs (on an average by 4% to 5%) in the lung relative to either of the 2 dose algorithms. Crown
Copyright © 2013 Published by American Association of Medical Dosimetrists on behalf of American Association of Medical Dosimetrists All rights reserved.

Entities:  

Keywords:  Acuros XB; Analytic anisotropic algorithm; Lung radiotherapy; SBRT planning

Mesh:

Year:  2013        PMID: 24333022     DOI: 10.1016/j.meddos.2013.10.003

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  3 in total

1.  Dosimetric effects of the acuros XB and anisotropic analytical algorithm on volumetric modulated arc therapy planning for prostate cancer using an endorectal balloon.

Authors:  Taeryool Koo; Jin-Beom Chung; Keun-Yong Eom; Jin-Yong Seok; In-Ah Kim; Jae-Sung Kim
Journal:  Radiat Oncol       Date:  2015-02-22       Impact factor: 3.481

2.  Verification of Acuros XB dose algorithm using 3D printed low-density phantoms for clinical photon beams.

Authors:  Rodolfo Zavan; Philip McGeachy; Joseph Madamesila; Jose-Eduardo Villarreal-Barajas; Rao Khan
Journal:  J Appl Clin Med Phys       Date:  2018-03-25       Impact factor: 2.102

3.  The dosimetric effect of electron density overrides in 3DCRT Lung SBRT for a range of lung tumor dimensions.

Authors:  Grace E A Healy; Steven H Marsh; Andrew T Cousins
Journal:  J Appl Clin Med Phys       Date:  2018-09-10       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.