Single center experience with the Sorin Bicarbon prosthesis: a 17-year clinical follow-up.

Michele Celiento1, Luca Filaferro1, Aldo D Milano2, Gerardo Anastasio1, Gabriele Ferrari1, Uberto Bortolotti3.   

Abstract

OBJECTIVE: To evaluate the long-term results of aortic valve replacement (AVR) and mitral valve replacement (MVR) with the Sorin Bicarbon prosthesis (SBP).
METHODS: Five hundred seven patients (306 men, 201 women), mean age 62±10 years (range, 21-86 years), received an SBP between 1994 and 2000; AVR was performed in 344 (67%) and MVR in 163 (33%). The main concomitant procedure was coronary artery grafting in 79 patients (16%). Follow-up was 99% complete; mean follow-up was 12.7±4.0 years with a cumulative duration of follow-up of 6475 patient-years in the entire group (4348 patient-years for AVR and 2124 patient-years for MVR).
RESULTS: Hospital mortality was 2.7% (AVR, 2.03%; MVR, 4.3%). There were 169 late deaths (AVR, 128; MVR, 41). Actuarial survival at 17 years is 49.7%±5.3% for AVR and 62.0%±6.1% for MVR. At the last follow-up, 310 survivors (199 AVR, 111 MVR) are in New York Heart Association functional class I or II. At 17 years, actuarial freedom from valve-related deaths, embolism, and bleeding is 89.8%±4.8%, 85.8%±5.4%, and 96.2%±1.2% after AVR, and 91.9%±3.9%, 96.3%±1.8%, 95.0%±2.9% after MVR. Reoperation was required in 5 patients with AVR (thrombosis in 4 and perivalvular leak in 1). Actuarial freedom from reoperation is 98.1%±0.8% after AVR and 100% after MVR; freedom from endocarditis is 100% after AVR and 99.2%±0.7% after MVR. No cases of intrinsic structural valve failure were observed.
CONCLUSIONS: The SBP has shown excellent results in terms of clinical improvement and freedom from valve-related complications, even up to 17 years after AVR and MVR. It therefore seems to be a safe option whenever a mechanical prosthesis is needed.
Copyright © 2014 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 24332112     DOI: 10.1016/j.jtcvs.2013.11.015

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  3 in total

Review 1.  Valve-related complications after mechanical heart valve implantation.

Authors:  Yoshio Misawa
Journal:  Surg Today       Date:  2014-12-18       Impact factor: 2.549

2.  Low-dose anticoagulation after isolated mechanical aortic valve replacement with Liva Nova Bicarbon prosthesis: A post hoc analysis of LOWERING-IT Trial.

Authors:  Michele Torella; Iolanda Aquila; Paolo Chiodini; Cristiano Amarelli; Gianpaolo Romano; Ester Elena Della Ratta; Marisa De Feo; Alessandro Della Corte; Ciro Indolfi; Daniele Torella; Luca Salvatore De Santo
Journal:  Sci Rep       Date:  2018-05-30       Impact factor: 4.379

3.  Nineteen Years of Single Institute Experiences with Sorin Bicarbon Prosthesis.

Authors:  Hung Dung Van
Journal:  Ann Thorac Cardiovasc Surg       Date:  2019-03-12       Impact factor: 1.520

  3 in total