Literature DB >> 24321494

Update of NIST half-life results corrected for ionization chamber source-holder instability.

M P Unterweger1, R Fitzgerald2.   

Abstract

As reported at the ICRM 2011, it was discovered that the source holder used for calibrations in the NIST 4πγ ionization chamber (IC) was not stable. This has affected a large number of half-life measurement results previously reported and used in compilations of nuclear data. Corrections have been made on all of the half-life data based on the assumption that the changes to the ionization chamber response were gradual. The corrections are energy dependent and therefore radionuclide specific. This presentation will review our results and present the recommended changes in half-life values and/or uncertainties.
© 2013 Published by Elsevier Ltd.

Entities:  

Keywords:  Calibrations; Half-life measurement; Ionization chamber

Year:  2013        PMID: 24321494      PMCID: PMC7649977          DOI: 10.1016/j.apradiso.2013.11.017

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  Half-life measurements at the National Institute of Standards and Technology.

Authors:  M P Unterweger
Journal:  Appl Radiat Isot       Date:  2002 Jan-Feb       Impact factor: 1.513

2.  Unexpected bias in NIST 4πγ ionization chamber measurements.

Authors:  M P Unterweger; R Fitzgerald
Journal:  Appl Radiat Isot       Date:  2012-04-04       Impact factor: 1.513

3.  NIST Ionization Chamber "A" Sample-Height Corrections.

Authors:  Ryan Fitzgerald
Journal:  J Res Natl Inst Stand Technol       Date:  2012-02-02
  3 in total
  2 in total

1.  Corrigendum to "Update of NIST half-life results corrected for ionization chamber source-holder instability" [Appl. Radiat. Isot. 87 (2014) 92-94].

Authors:  M P Unterweger; R Fitzgerald
Journal:  Appl Radiat Isot       Date:  2020-02-07       Impact factor: 1.513

2.  Tandem Mass Spectrometry as an Independent Method for Corroborating Fluorine-18 Radioactivity Measurements in Positron Emission Tomography.

Authors:  H Umesha Shetty; Cheryl L Morse; Victor W Pike
Journal:  ACS Meas Sci Au       Date:  2022-06-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.