Characteristics of flattening filter free beams at low monitor unit settings.

Yuichi Akino1, Seiichi Ota, Shinichi Inoue, Hirokazu Mizuno, Iori Sumida, Yasuo Yoshioka, Fumiaki Isohashi, Kazuhiko Ogawa.   

Abstract

PURPOSE: Newer linear accelerators (linacs) have been equipped to deliver flattening filter free (FFF) beams. When FFF beams are used for step-and-shoot intensity-modulated radiotherapy (IMRT), the stability of delivery of small numbers of monitor units (MU) is important. The authors developed automatic measurement techniques to evaluate the stability of the dose profile, dose linearity, and consistency. Here, the authors report the performance of the Artiste™ accelerator (Siemens, Erlangen, Germany) in delivering low-MU FFF beams.
METHODS: A 6 MV flattened beam (6X) with 300 MU/min dose rate and FFF beams of 7 (7XU) and 11 MV (11XU), each with a 500 MU/min dose rate, were measured at 1, 2, 3, 5, 8, 10, and 20 MU settings. For the 2000 MU/min dose rate, the 7 (7XUH) and 11 MV (11XUH) beams were set at 10, 15, 20, 25, and 30 MU because of the limits of the minimum MU settings. Beams with 20 × 20 and 10 × 10 cm(2) field sizes were alternately measured ten times in intensity modulated (IM) mode, with which Siemens linacs regulate beam delivery for step-and-shoot IMRT. The in- and crossplane beam profiles were measured using a Profiler™ Model 1170 (Sun Nuclear Corporation, Melbourne, FL) in multiframe mode. The frames of 20 × 20 cm(2) beams were identified at the off-axis profile. The 6X beam profile was normalized at the central axis. The 7 and 11 MV FFF beam profiles were rescaled to set the dose at the central axis at 145% and 170%, respectively. Point doses were also measured using a Farmer-type ionization chamber and water-equivalent solid phantom to evaluate the linearity and consistency of low-MU beam delivery. The values displayed on the electrometer were recognized with a USB-type camera and read with open-source optical character recognition software.
RESULTS: The symmetry measurements of the 6X, 7XU, and 11XU beam profiles were better than 2% for beams ≥ 2 MU and improved with increasing MU. The variations in flatness of FFF beams ≥ 2 MU were ± 5%. The standard deviation of the symmetry and flatness also decreased with increasing MU. The linearity of the 6X beam was ± 1% and ± 2% for the beams of ≥ 5 and ≥ 3 MU, respectively. The 7XU and 11XU beams of ≥ 2 MU showed linearity with ± 2% except the 7XU beam of 8 MU (+2.9%). The profiles of the FFF beams with 2000 and 500 MU/min dose rate were similar.
CONCLUSIONS: The characteristics of low-MU beams delivered in IM mode were evaluated using an automatic measurement system developed in this study. The authors demonstrated that the profiles of FFF beams of the Artiste™ linac were highly stable, even at low MU. The linearity of dose output was also stable for beams ≥ 2 MU.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320454     DOI: 10.1118/1.4824920

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Analyzing the characteristics of 6 MV photon beam at low monitor unit settings.

Authors:  L Nithya; N Arunai Nambi Raj; Sasikumar Rathinamuthu
Journal:  J Med Phys       Date:  2016 Jan-Mar

2.  Monte-Carlo simulation of the Siemens Artiste linear accelerator flat 6 MV and flattening-filter-free 7 MV beam line.

Authors:  Alemeh Sadrollahi; Frank Nuesken; Norbert Licht; Christian Rübe; Yvonne Dzierma
Journal:  PLoS One       Date:  2019-01-08       Impact factor: 3.240

3.  Simultaneous integrated boost (SIB) radiation therapy of right sided breast cancer with and without flattening filter - A treatment planning study.

Authors:  Johannes Maier; Bernadette Knott; Manuel Maerz; Rainer Loeschel; Oliver Koelbl; Barbara Dobler
Journal:  Radiat Oncol       Date:  2016-08-31       Impact factor: 3.481

4.  Re-irradiating spinal column metastases using IMRT and VMAT with and without flattening filter - a treatment planning study.

Authors:  Barbara Dobler; Amine Khemissi; Tina Obermeier; Matthias G Hautmann; Zaira Katsilieri; Oliver Kölbl
Journal:  Radiat Oncol       Date:  2016-03-01       Impact factor: 3.481

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.