Evaluation of an aSi-EPID with flattening filter free beams: applicability to the GLAaS algorithm for portal dosimetry and first experience for pretreatment QA of RapidArc.

G Nicolini1, A Clivio, E Vanetti, H Krauss, P Fenoglietto, L Cozzi, A Fogliata.   

Abstract

PURPOSE: To demonstrate the feasibility of portal dosimetry with an amorphous silicon mega voltage imager for flattening filter free (FFF) photon beams by means of the GLAaS methodology and to validate it for pretreatment quality assurance of volumetric modulated arc therapy (RapidArc).
METHODS: The GLAaS algorithm, developed for flattened beams, was applied to FFF beams of nominal energy of 6 and 10 MV generated by a Varian TrueBeam (TB). The amorphous silicon electronic portal imager [named mega voltage imager (MVI) on TB] was used to generate integrated images that were converted into matrices of absorbed dose to water. To enable GLAaS use under the increased dose-per-pulse and dose-rate conditions of the FFF beams, new operational source-detector-distance (SDD) was identified to solve detector saturation issues. Empirical corrections were defined to account for the shape of the profiles of the FFF beams to expand the original methodology of beam profile and arm backscattering correction. GLAaS for FFF beams was validated on pretreatment verification of RapidArc plans for three different TB linacs. In addition, the first pretreatment results from clinical experience on 74 arcs were reported in terms of γ analysis.
RESULTS: MVI saturates at 100 cm SDD for FFF beams but this can be avoided if images are acquired at 150 cm for all nominal dose rates of FFF beams. Rotational stability of the gantry-imager system was tested and resulted in a minimal apparent imager displacement during rotation of 0.2 ± 0.2 mm at SDD = 150 cm. The accuracy of this approach was tested with three different Varian TrueBeam linacs from different institutes. Data were stratified per energy and machine and showed no dependence with beam quality and MLC model. The results from clinical pretreatment quality assurance, provided a gamma agreement index (GAI) in the field area for six and ten FFF beams of (99.8 ± 0.3)% and (99.5 ± 0.6)% with distance to agreement and dose difference criteria set to 3 mm/3% with 2 mm/2% thresholds, GAI resulted (95.7.0 ± 2.3)% and (97.2 ± 2.1)%.
CONCLUSIONS: The GLAaS methodology, introduced in clinical practice for conventional flattened photon beams for machine, IMRT, and RapidArc quality assurance, was successfully adapted for FFF beams of Varian TrueBeam Linac. The detector saturation effects could be eliminated if the portal images acquired at 150 cm for all nominal dose rates of FFF beams.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320427     DOI: 10.1118/1.4824923

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  SBRT/SRS patient-specific QA using GAFchromicTM EBT3 and FilmQATM Pro software.

Authors:  Giuseppe Stella; Nina Cavalli; Elisa Bonanno; Lucia Zirone; Giuseppina Rita Borzì; Martina Pace; Andrea Girlando; Anna M Gueli; Carmelo Marino
Journal:  J Radiosurg SBRT       Date:  2022

2.  Evaluation of the Machine Performance Check application for TrueBeam Linac.

Authors:  Alessandro Clivio; Eugenio Vanetti; Steven Rose; Giorgia Nicolini; Maria F Belosi; Luca Cozzi; Christof Baltes; Antonella Fogliata
Journal:  Radiat Oncol       Date:  2015-04-21       Impact factor: 3.481

3.  EPID-based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams.

Authors:  Narges Miri; Peter Keller; Benjamin J Zwan; Peter Greer
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

4.  Practical approach for pretreatment verification of IMRT with flattening filter free(FFF) beams using Varian Portal Dosimetry.

Authors:  Soonki Min; Young Eun Choi; Jungwon Kwak; Byungchul Cho
Journal:  J Appl Clin Med Phys       Date:  2014-01-08       Impact factor: 2.102

5.  Pixel response-based EPID dosimetry for patient specific QA.

Authors:  Bin Han; Aiping Ding; Minghui Lu; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-12-15       Impact factor: 2.102

6.  Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams.

Authors:  Zhigang Xu; Jinkoo Kim; James Han; An Ting Hsia; Samuel Ryu
Journal:  J Appl Clin Med Phys       Date:  2018-05-21       Impact factor: 2.102

7.  Dosimetric Validation of Digital Megavolt Imager for Flattening Filter Free Beams in the Pre-Treatment Quality Assurance of Stereotactic Body Radiation Therapy for Liver Metastases.

Authors:  Vendhan Subramani; Murali Rathakrishnan; Arunai Nambiraj N; Saraswathi Chitra S; Murali Venkatraman
Journal:  Asian Pac J Cancer Prev       Date:  2020-06-01

8.  On flattening filter-free portal dosimetry.

Authors:  Eduardo Pardo; Juan Castro Novais; María Yolanda Molina López; Sheila Ruiz Maqueda
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.