Mechanical resistance of screwless morse taper and screw-retained implant-abutment connections.

Cihan Sadi Ugurel1, Martin Steiner, Gul Isik-Ozkol, Omer Kutay, Matthias Kern.   

Abstract

OBJECTIVE: The screwless morse taper (SMT) implant-abutment connection is an alternative to conventional external or internal screw-retained (ISR) connections. The aim of this in vitro study was to evaluate mechanical resistance of the SMT connection and to compare it with three different ISR connections.
MATERIALS AND METHODS: Four implant systems were tested in this study: SMT system; Tasarimmed Octo (Istanbul, Turkey), ISR systems; Straumann Bone Level (Basel, Switzerland), Biohorizons Internal (Birmingham, AL, USA), and Dentsply Friadent Xive (Mannheim, Germany). Overall, 64 specimens with stylized single crowns were prepared: 32 for dynamic loading (DL) and 32 for static loading (SL). DL was carried out using a chewing simulator with 120 N at 1.75 Hz for 1.2 × 10(6) cycles, and SL was performed with a universal testing machine at a crosshead speed of 2 mm/min with an angulation of 30°. Cycles until failure from DL and fracture/bending loads at SL were recorded. Statistical analyses were made with Dunn's multiple comparison.
RESULTS: Median cycles until failure in DL were as follows: Octo 86,354 (24,810-153,875), Straumann 1,200,000 (1,156,618-1,200,000), Biohorizons 539,719 (437,224-858,732), Xive 139,411 (139,411-139,411). Median fracture/bending loads in Newton at SL were as follows: Octo 429.6 (404.5-482.7), Straumann 574.6 (544.6-629.9), Biohorizons 548.7 (532.9-567.3), Xive 431.5 (412.5-520.5). There were significant differences between the implant systems under both loading conditions (P ≤ 0.05) revealing that the Octo implant system's SMT connection showed significantly lower cycles to failures and lower fracture/bending loads compared with the ISR connections of the Straumann and Biohorizons implant systems. However, there was no significant difference compared with the Xive implant system.
CONCLUSION: The mechanical resistance of the screwless morse taper implant system is lower than that of the ISR implant systems that might result in more frequent clinical complications.
© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  dental implant-abutment connection; mechanical strength; morse taper

Mesh:

Year:  2013        PMID: 24313278     DOI: 10.1111/clr.12303

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  5 in total

1.  The Mechanical Behavior of a Screwless Morse Taper Implant-Abutment Connection: An In Vitro Study.

Authors:  Aimen Bagegni; Vincent Weihrauch; Kirstin Vach; Ralf Kohal
Journal:  Materials (Basel)       Date:  2022-05-08       Impact factor: 3.748

2.  A New Model to Study Fatigue in Dental Implants Based on Probabilistic Finite Elements and Cumulative Damage Model.

Authors:  María Prados-Privado; José Antonio Bea; Rosa Rojo; Sérgio A Gehrke; José Luis Calvo-Guirado; Juan Carlos Prados-Frutos
Journal:  Appl Bionics Biomech       Date:  2017-07-05       Impact factor: 1.781

3.  Effect of implant placement depth on the peri-implant bone defect configurations in ligature-induced peri-implantitis: An experimental study in dogs.

Authors:  B Huang; L Zhang; L Xu; W Zhu; L Witek; N Tovar; P-G Coelho; H Meng
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2018-01-01

4.  Effect of the prosthetic index on stress distribution in Morse taper connection implant system and peri-implant bone: a 3D finite element analysis.

Authors:  Wen-Tao Zhang; Kang-Jie Cheng; Fan Yang; Lin-Hong Wang; Yun-Feng Liu; Russell Wang; Yun-Fang Chen; Yu-de Ding
Journal:  BMC Oral Health       Date:  2022-09-30       Impact factor: 3.747

5.  Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study.

Authors:  Sung-Jun Kim; KeunBaDa Son; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2018-10-22       Impact factor: 1.904

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.