Assessment of central blood pressure in patients with type 2 diabetes: a comparison between SphygmoCor and invasively measured values.

Esben Laugesen1, Niklas B Rossen, Christian D Peters, Michael Mæng, Eva Ebbehøj, Søren T Knudsen, Klavs W Hansen, Hans E Bøtker, Per L Poulsen.   

Abstract

BACKGROUND: The SphygmoCor is used for noninvasive assessment of ascending aortic blood pressure (BP). However, the validity of the SphygmoCor transfer function has not been tested in an exclusively type 2 diabetic patient sample. Calibration with systolic (SBP) and diastolic (DBP) brachial BP has previously been associated with substantial imprecision of central BP estimates. We hypothesized that different noninvasive calibration strategies might improve the accuracy of the estimated ascending aortic BPs.
METHODS: In 34 patients with type 2 diabetes we estimated ascending aortic SBP and DBP using the SphygmoCor device and compared these data with invasively recorded data. The validity of the transfer function was assessed by calibrating with invasively recorded DBP and mean BP (MBP). The influence of noninvasive calibration strategies was assessed by calibrating with brachial oscillometric SBP+DBP vs. DBP+MBP using a form factor (ff) of 0.33 and 0.40, respectively.
RESULTS: When calibrating with invasive BP, the difference between estimated and invasively measured ascending aortic SBP and DBP was -2.3±5.6/1.0±0.9 mm Hg. When calibrating with oscillometric brachial BPs, the differences were -9.6±8.1/14.1±6.2 mm Hg (calibration with SBP and DBP), -8.3±11.7/13.9±6.1 mm Hg (DBP and MBP; ff = 0.33), and 1.9±12.2/14.1±6.2 mm Hg (DBP and MBP; ff = 0.40), respectively. Calibration with the average of 3 brachial BPs did not improve accuracy.
CONCLUSIONS: The SphygmoCor transfer function seems valid in patients with type 2 diabetes. Noninvasive calibration with DBP and MBP (ff = 0.40) enables accurate estimation of mean ascending aortic SBP at the group level. However, the wide limits of agreement indicate limited accuracy in the individual patient. CLINICAL TRIALS REGISTRATION: Clinical Trials No. NCT01538290.

Entities:  

Keywords:  applanation tonometry; blood pressure; blood pressure monitoring; central blood pressure; hypertension; invasive blood pressure monitoring; transfer function; type 2 diabetes; validation.

Mesh:

Year:  2013        PMID: 24304654     DOI: 10.1093/ajh/hpt195

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  5 in total

1.  Mean arterial pressure values calculated using seven different methods and their associations with target organ deterioration in a single-center study of 1878 individuals.

Authors:  Theodore G Papaioannou; Athanase D Protogerou; Dimitrios Vrachatis; Giorgos Konstantonis; Evaggelia Aissopou; Antonis Argyris; Efthimia Nasothimiou; Elias J Gialafos; Marianna Karamanou; Dimitris Tousoulis; Petros P Sfikakis
Journal:  Hypertens Res       Date:  2016-05-19       Impact factor: 3.872

2.  Practical Suitability of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Review of the Microlife WatchBP Office Central.

Authors:  Willem J Verberk; Hao-Min Cheng; Li-Chih Huang; Chia-Ming Lin; Yao-Pin Teng; Chen-Huan Chen
Journal:  Pulse (Basel)       Date:  2016-02-12

3.  Transfer-function-free technique for the noninvasive determination of the human arterial pressure waveform.

Authors:  Alessandro Giudici; Carlo Palombo; Carmela Morizzo; Michaela Kozakova; J Kennedy Cruickshank; Ian B Wilkinson; Ashraf W Khir
Journal:  Physiol Rep       Date:  2021-09

Review 4.  Aortic stiffness: pathophysiology, clinical implications, and approach to treatment.

Authors:  Salil Sethi; Oscar Rivera; Rene Oliveros; Robert Chilton
Journal:  Integr Blood Press Control       Date:  2014-05-23

5.  Validation of BP devices QardioArm® in the general population and Omron M6 Comfort® in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP).

Authors:  Mirna N Chahine; Jirar Topouchian; Parounak Zelveian; Zoya Hakobyan; Arevik Melkonyan; Alaa Azaki; Reem Diab; Aya Harb; Roland Asmar
Journal:  Med Devices (Auckl)       Date:  2017-12-27
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.