Literature DB >> 24304290

Pre-fabricated zirconium dioxide implant abutments for single-tooth replacement in the posterior region: success and failure after 3 years of function.

Frank P Nothdurft1, Joerg Nonhoff, Peter R Pospiech.   

Abstract

OBJECTIVE: Zirconia implant abutments have gained a much broader clinical use over the past few years. The aim of the present study was to assess the clinical performance of a pre-fabricated zirconium dioxide implant abutment for single-tooth replacement in the posterior region.
MATERIALS AND METHODS: Forty implants of the XiVE(®) S plus screw type (DENTSPLY Friadent, Mannheim, Germany) were inserted in the posterior region of 24 patients and provided with zirconium dioxide abutments (FRIADENT(®) CERCON(®) Abutment, DENTSPLY Friadent). The following parameters were used to document the state of soft tissue: modified plaque index, modified sulcus bleeding index and pocket depth. Mesial and distal bone levels were determined on radiographs during the prosthetic treatment and at the 36-month recall.
RESULTS: Thirty-seven implants could be followed up after 36 months in function. One patient wearing two abutments was lost to follow-up. One abutment exhibited a rotational misfit after 2 years in function. A further abutment showed the same failure at the 36-months recall appointment. In the remaining 36 implants the soft and hard tissue parameters were indicative of a low inflammatory status. Compared to the baseline situation, a partly significant bone apposition could be observed. Chipping of parts of the veneering ceramic was registered in 22% of the remaining implant restorations.
CONCLUSIONS: The use of zirconia abutments in this study lead to mainly healthy peri-implant hard and soft tissue conditions but, considering the observed failures after 3 years in function, clinical long-term results should be awaited before recommending full zirconia implant abutments in a posterior indication.

Entities:  

Keywords:  Biomaterials; clinical research; clinical trials; soft tissue-implant interactions; zirconia

Mesh:

Substances:

Year:  2013        PMID: 24304290     DOI: 10.3109/00016357.2013.863970

Source DB:  PubMed          Journal:  Acta Odontol Scand        ISSN: 0001-6357            Impact factor:   2.331


  4 in total

1.  Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing.

Authors:  Dieter Pils; Ralf J Baeppler; Rüdiger Junker; Andrej M Kielbassa; Frank P Nothdurft
Journal:  Clin Oral Investig       Date:  2019-05-10       Impact factor: 3.573

2.  The prognosis of splinted restoration of the most-distal implants in the posterior region.

Authors:  Jong-Bin Lee; Man-Yong Kim; Chang-Sung Kim; Young-Taek Kim
Journal:  J Adv Prosthodont       Date:  2016-12-15       Impact factor: 1.904

3.  All-ceramic versus titanium-based implant supported restorations: Preliminary 12-months results from a randomized controlled trial.

Authors:  Paul Weigl; Georgia Trimpou; Eleftherios Grizas; Pablo Hess; Georg-Hubertus Nentwig; Hans-Christoph Lauer; Jonas Lorenz
Journal:  J Adv Prosthodont       Date:  2019-02-26       Impact factor: 1.904

Review 4.  A systematic review and meta-analysis evaluating the survival, the failure, and the complication rates of veneered and monolithic all-ceramic implant-supported single crowns.

Authors:  Bjarni Elvar Pjetursson; Irena Sailer; Andrey Latyshev; Kerstin Rabel; Ralf-Joachim Kohal; Duygu Karasan
Journal:  Clin Oral Implants Res       Date:  2021-10       Impact factor: 5.021