Literature DB >> 24304098

Novel flow cytometric screening method for bacterial contamination of red blood cells: a proof-of-principle evaluation.

Tanja Vollmer1, Cornelius Knabbe, Jens Dreier.   

Abstract

BACKGROUND: Platelet concentrates (PCs) are the major source of transfusion-transmitted bacterial infection but the significantly higher number of transfused red blood cells (RBCs) has resulted in a similar occurrence of possible transfusion-associated bacterial infections as those of recipients transfused with platelets. The aim of this study was adaption of the BactiFlow (BF) flow cytometric rapid PC screening method for the detection of bacterial contamination in RBCs. STUDY DESIGN AND METHODS: The BF assay was used to detect and count bacteria based on esterase activity in viable cells in RBCs. The initial sample preparation for RBCs included the lysis of RBCs, followed by enrichment of bacteria by centrifugation and subsequent standard PC sample preparation. Two-hundred RBC units were analyzed by BF and the BacT/Alert automated culture system. Furthermore, RBCs were spiked with eight different bacteria to monitor bacterial growth kinetics.
RESULTS: The BF showed an excellent correlation to conventional plate count results after RBC sample processing. The diagnostic sensitivity of the assay was determined to fewer than 500 counts/mL. Analysis of RBC units showed concordant negative results by BF and culture. Growth kinetics of bacteria were successfully monitored using flow cytometry, showing that Yersinia enterocolitica or Serratia liquefaciens had the capability to grow in RBCs to high counts.
CONCLUSION: Our study demonstrates the successful modification and application of a rapid flow cytometric screening method for bacterial contamination in RBCs. The BF assay represents a powerful basic tool to study bacterial contamination, potential screening strategies of RBCs, and the clinical outcome of screened RBCs prepared for transfusion.
© 2013 AABB.

Entities:  

Mesh:

Year:  2013        PMID: 24304098     DOI: 10.1111/trf.12513

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  2 in total

1.  Bench Test for the Detection of Bacterial Contamination in Platelet Concentrates Using Rapid and Cultural Detection Methods with a Standardized Proficiency Panel.

Authors:  Tanja Vollmer; Cornelius Knabbe; Wolf-Jochen Geilenkeuser; Michael Schmidt; Jens Dreier
Journal:  Transfus Med Hemother       Date:  2015-07-16       Impact factor: 3.747

2.  Multimodal Magneto-Fluorescent Nanosensor for Rapid and Specific Detection of Blood-Borne Pathogens.

Authors:  Tuhina Banerjee; Tanuja Tummala; Rebekah Elliott; Vedant Jain; Wesley Brantley; Laci Hadorn; Santimukul Santra
Journal:  ACS Appl Nano Mater       Date:  2019-08-19