Metaphyseal screw augmentation of the LISS-PLT plate with polymethylmethacrylate improves angular stability in osteoporotic proximal third tibial fractures: a biomechanical study in human cadaveric tibiae.

Michael Goetzen1, Tomas Nicolino, Ladina Hofmann-Fliri, Michael Blauth, Markus Windolf.   

Abstract

OBJECTIVES: The incidence of osteoporotic proximal tibial fractures has increased during the last 2 decades. A promising approach in osteoporotic fracture fixation is polymethylmethacrylate-based cement augmentation of implants to gain better implant purchase in the bone. This study investigates the biomechanical benefits of screw augmentation in less invasive stabilization system-proximal lateral tibial (LISS-PLT) plates in cadaveric extraarticular comminuted proximal tibial fractures (OTA-41-A3.3).
METHODS: Standardized extraarticular proximal tibial fractures were stabilized with the LISS-PLT plate in 6 paired osteoporotic cadaveric tibiae. Bone mineral density was measured with high-resolution, quantitative computed tomography scans to identify bone quality. In the augmented group, the 5 proximal screws of the LISS-PLT plate were augmented with 1 mL of bone cement each, whereas the contralateral tibia was instrumented conventionally as the control. Cyclic axial loading was applied to each specimen with a starting load of 150 N, using a ramp of 0.05 N per cycle to 10-mm axial displacement. Varus displacement was identified from anterior-posterior radiographs.
RESULTS: Bone mineral density showed no significant difference between the 2 groups (P = 0.47). The nonaugmented group reached 9417 load cycles (SD 753) until failure, compared with 14,792 load cycles (SD 2088) in the augmented group (P = 0.002). In the early-onset failure (deformation at 8250 load cycles), varus displacement was significantly smaller in the augmented group (0.46 degrees, SD 0.6) than in the nonaugmented group (3.23 degrees, SD 1.7) (P = 0.01).
CONCLUSIONS: This biomechanical study showed that cement augmentation of the LISS-PLT plate screws in osteoporotic proximal extraarticular tibial fractures significantly lowers the propensity toward screw migration and secondary varus displacement.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24296600     DOI: 10.1097/BOT.0000000000000004

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  Does metaphyseal cement augmentation in fracture management influence the adjacent subchondral bone and joint cartilage?: an in vivo study in sheep stifle joints.

Authors:  Michael Goetzen; Ladina Hofmann-Fliri; Daniel Arens; Stephan Zeiter; Vincent Stadelmann; Dirk Nehrbass; R Geoff Richards; Michael Blauth
Journal:  Medicine (Baltimore)       Date:  2015-01       Impact factor: 1.889

2.  The Photodynamic Bone Stabilization System: a minimally invasive, percutaneous intramedullary polymeric osteosynthesis for simple and complex long bone fractures.

Authors:  Paul Vegt; Jeffrey M Muir; Jon E Block
Journal:  Med Devices (Auckl)       Date:  2014-12-12

3.  Effect of the lipoxygenase inhibitor baicalein on bone tissue and bone healing in ovariectomized rats.

Authors:  Marina Komrakova; Stephan Sehmisch; Dominik Saul; Marie Weber; Marc Hendrik Zimmermann; Robyn Laura Kosinsky; Daniel Bernd Hoffmann; Björn Menger; Stefan Taudien; Wolfgang Lehmann
Journal:  Nutr Metab (Lond)       Date:  2019-01-11       Impact factor: 4.169

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.