Literature DB >> 24286940

Evaluating patency rates of an ultralow-porosity expanded polytetrafluoroethylene covered stent in the treatment of venous stenosis in arteriovenous dialysis circuits.

Michael G Chan1, Franklin J Miller2, Karim Valji3, Anshuman Bansal4, Michael D Kuo5.   

Abstract

PURPOSE: To evaluate the efficacy of an ultralow-porosity expanded polytetrafluoroethylene (ePTFE) covered stent in the treatment of autogenous arteriovenous fistula (AVF) and prosthetic arteriovenous graft (AVG) venous outflow stenoses.
MATERIALS AND METHODS: Clinical and angiographic outcomes of 20 consecutive patients with arteriovenous dialysis circuits treated with the endoprosthesis were reviewed following institutional review board approval. Patients were followed routinely at 2 months and 6 months after stent placement, or earlier if clinically warranted. The primary endpoint was 2- and 6-month primary treatment area patency. Secondary endpoints included primary circuit patency, primary assisted patency, and secondary patency.
RESULTS: Eleven patients with AVFs and nine patients with AVGs were treated successfully with the covered stent. Primary treatment area patency rates were 85% ± 16 at both 2 months and 6 months. Primary circuit patency rates were was 65% ± 21 and 45% ± 22, respectively; primary assisted patency rates were 90% ± 13 and 85% ± 16, respectively; and secondary patency rates were 100% and 90% ± 13, respectively. Of the three cases of lost primary treatment area patency, two developed thrombosis and one developed recurrent stenosis. No significant differences were found between patients with AVFs and AVGs.
CONCLUSIONS: Data from this preliminary study suggests that the ultralow-porosity ePTFE covered stent may be a clinically viable option for treatment of venous outflow stenoses in arteriovenous vascular access circuits.
© 2014 SIR Published by SIR All rights reserved.

Entities:  

Keywords:  AVF; AVG; FDA; Food and Drug Administration; VAC; arteriovenous fistula; arteriovenous graft; ePTFE; expanded polytetrafluoroethylene; vascular access circuit

Mesh:

Substances:

Year:  2013        PMID: 24286940     DOI: 10.1016/j.jvir.2013.10.006

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  4 in total

Review 1.  Stent and Stent-Graft Use in Arteriovenous Dialysis Access.

Authors:  Gordon McLennan
Journal:  Semin Intervent Radiol       Date:  2016-03       Impact factor: 1.513

Review 2.  Important Contributions to the Interventional Radiology Dialysis Literature, 2014-2015.

Authors:  Charles E Ray; Brian Funaki; Ron C Gaba
Journal:  Semin Intervent Radiol       Date:  2016-03       Impact factor: 1.513

3.  Treatment of failing arterio-venous dialysis graft by angioplasty, stent, and stent graft: Two-years analysis of patency rates and cost-effectiveness.

Authors:  Jan Kavan; Jaroslav Kudlicka; Jan Malik; Eva Chytilova; Lukas Lambert; Marcela Slavikova; Patrik Matras; Andrea Burgetova
Journal:  Exp Ther Med       Date:  2019-09-25       Impact factor: 2.447

4.  An atypical case of hemodialysis access stent migration.

Authors:  Juan C Duque; Marwan Tabbara; Laisel Martinez; Karen Manzur-Pineda; Roberto I Vazquez-Padron; Adriana Dejman
Journal:  Clin Nephrol Case Stud       Date:  2022-01-24