Application of real-time ultrasound elastography for differential diagnosis of breast tumors.

Xia Gong1, Yi Wang, Ping Xu.   

Abstract

OBJECTIVES: The purpose of this study was to explore the clinical value of real-time ultrasound elastography in differentiating malignant from benign breast tumors and to determine an optimal cutoff for the traced area ratio by receiver operator characteristic (ROC) analysis for differential diagnosis between malignant and benign breast masses.
METHODS: From October 2010 to June 2011, 102 patients with 192 breast tumors were enrolled. Conventional sonograms and real-time elastograms were obtained from the patients. The sensitivity, specificity, and accuracy rates for sonography and elastography were calculated, and an ROC analysis was performed.
RESULTS: Tumors with an elasticity grade of 4 or higher were defined as malignant. The sensitivity, specificity, and accuracy of elastography were 92.65%, 73.39%, and 81.25%, respectively. These values were similar to those for conventional sonography. However, when the techniques were combined, the sensitivity, specificity, and accuracy increased to 88.23%, 95.97%, and 93.23%. A value of 1.65 was determined to be the traced area ratio cutoff by the ROC analysis and was used in this study for differential diagnosis. The sensitivity, specificity, and accuracy of this value were 76.47%, 96.77%, and 89.58%, respectively.
CONCLUSIONS: Both conventional sonography and ultrasound elastography could be used to differentiate malignant and benign breast tumors. If these techniques were combined, the diagnostic values would improve. In addition, a traced area ratio of 1.65 could be used as a cutoff to differentiate benign and malignant breast tumors.

Entities:  

Keywords:  breast tumor; conventional sonography; real-time elastography; traced area ratio

Mesh:

Year:  2013        PMID: 24277900     DOI: 10.7863/ultra.32.12.2171

Source DB:  PubMed          Journal:  J Ultrasound Med        ISSN: 0278-4297            Impact factor:   2.153


  3 in total

1.  Ultrasound elastography and magnetic resonance examinations are effective for the accurate diagnosis of mammary duct ectasia.

Authors:  Feixue Zhang; Dexin Yu; Mingming Guo; Qing Wang; Zhigang Yu; Fei Zhou; Meng Zhao; Feng Xue; Guangrui Shao
Journal:  Int J Clin Exp Med       Date:  2015-06-15

2.  Diagnostic evaluations of ultrasound and magnetic resonance imaging in mammary duct ectasia and breast cancer.

Authors:  Lei Song; Liang Li; Bin Liu; Dexin Yu; Fengguo Sun; Mingming Guo; Zhengmin Ruan; Feixue Zhang
Journal:  Oncol Lett       Date:  2017-12-05       Impact factor: 2.967

3.  Determining the elastography strain ratio cut off value for differentiating benign from malignant breast lesions: systematic review and meta-analysis.

Authors:  Timothy Musila Mutala; Gladys N Mwango; Angeline Aywak; Dania Cioni; Emanuele Neri
Journal:  Cancer Imaging       Date:  2022-02-12       Impact factor: 3.909

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.