Auricular burns associated with operating microscope use during otologic surgery.

Richard F Latuska1, Matthew L Carlson, Brian A Neff, Colin L Driscoll, George B Wanna, David S Haynes.   

Abstract

OBJECTIVE: To raise awareness of the potential hazard of auricular burns associated with operating microscope use during otologic surgery. STUDY
DESIGN: Retrospective case series and summary of the Food and Drug Administration's (FDA) Manufacturer and User Facility Device Experience (MAUDE) database of voluntary adverse event reports pertaining to microscope related auricular thermal injuries. PATIENTS AND
SETTING: All patients who sustained auricular burns while using the operating microscope during otologic surgery at 2 tertiary academic referral centers. MAIN OUTCOME MEASURES: Surgical procedure, microscope model, intensity of illumination, length of procedure, focal length, location and severity of burn, and patient outcome.
RESULTS: A total of 4 microscope-related auricular thermal injuries were identified from the authors' institutions. Additionally, 82 unique cases of soft tissue burns associated with the use of an operative microscope have been voluntarily reported to the FDA since 2004. A disproportionately large percent (∼ 30%) of these occurred within the field of otology, the majority of which were during tympanoplasty or tympanomastoidectomy procedures at focal length distances of 300 mm or less with xenon light source microscopes.
CONCLUSION: Simultaneous advancements in light delivery technologies and lens optics have continued to improve the efficiency of the operating microscope; however, these improvements also increase the potential for thermal injuries. Although rare, a review of the FDA MAUDE database suggests that microscope-related soft tissue burns occur more frequently in otology than any other surgical specialty. A variety of factors may help explain this finding, including the unique anatomy of the external ear with thin skin and limited underlying adipose tissue. Preventative measures should be taken to decrease the risk of thermal injuries including use of the lowest comfortable light intensity, adjusting the aperture width to match the operative field, frequent wound irrigation, and covering exposed portions of the pinna with a moist surgical sponge.

Entities:  

Mesh:

Year:  2014        PMID: 24270729     DOI: 10.1097/MAO.0b013e3182a5d340

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  4 in total

1.  Comprehensive review of surgical microscopes: technology development and medical applications.

Authors:  Ling Ma; Baowei Fei
Journal:  J Biomed Opt       Date:  2021-01       Impact factor: 3.170

2.  Intraoperative risk factors for peritumoral infarctions following glioma surgery.

Authors:  Per Sveino Strand; Sasha Gulati; Lisa Millgård Sagberg; Ole Solheim
Journal:  Brain Spine       Date:  2022-06-07

3.  Utility and safety of a novel surgical microscope laser light source.

Authors:  Taku Sato; Mudathir S Bakhit; Kyouichi Suzuki; Jun Sakuma; Masazumi Fujii; Yuta Murakami; Yuhei Ito; Tetsuo Sugano; Kiyoshi Saito
Journal:  PLoS One       Date:  2018-02-01       Impact factor: 3.240

4.  A Primer to the Structure, Content and Linkage of the FDA's Manufacturer and User Facility Device Experience (MAUDE) Files.

Authors:  Lisa Garnsey Ensign; K Bretonnel Cohen
Journal:  EGEMS (Wash DC)       Date:  2017-06-14
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.