In vivo longitudinal chromatic aberration of pseudophakic eyes.

Damian Siedlecki1, Agnieszka Jóźwik, Marek Zając, Aneta Hill-Bator, Anna Turno-Kręcicka.   

Abstract

PURPOSE: To present the results of longitudinal chromatic aberration measurements on two groups of pseudophakic eyes in comparison to healthy eyes.
METHODS: The longitudinal chromatic aberration of the eye, defined as chromatic difference of refraction with disabled accommodation, was measured with the use of a visual refractometer with a custom-designed target illuminator consisting of a narrow-band RGB diode (blue λb = 470 ± 15 nm; green λg = 525 ± 18 nm; red λr = 660 ± 10 nm). The measurements were performed on nine eyes implanted with AcrySof IQ SN60WF, 14 eyes implanted with AcrySof SA60AT, and 10 phakic eyes under cycloplegia.
RESULTS: The mean values of the longitudinal chromatic aberration between 470 and 660 nm for the control group was 1.12 ± 0.14 D. For SA60AT group, it was 1.45 ± 0.42 D whereas for SN60WF it was 1.17 ± 0.52 D. The statistical test showed significant difference between SA60AT and the control group (p << 0.05) and no significant difference between SN60WF and the control groups (p = 0.64).
CONCLUSIONS: The study showed that the longitudinal chromatic aberration in vivo can be easily and reliably estimated with an adapted visual refractometer. The two groups of pseudophakic eyes measured in this study showed different values of chromatic aberration. Its magnitude for SA60AT group was significantly larger than for the control group whereas for SN60WF the difference was not significant. The optical material used for intraocular lens design may have significant influence on the magnitude of the chromatic aberration of the pseudophakic eye, and therefore on its optical and visual performance in polychromatic light.

Entities:  

Mesh:

Year:  2014        PMID: 24270638     DOI: 10.1097/OPX.0000000000000137

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  3 in total

1.  Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models.

Authors:  Hui Song; Xiaoyong Yuan; Xin Tang
Journal:  BMC Ophthalmol       Date:  2016-01-11       Impact factor: 2.209

2.  Differences of Longitudinal Chromatic Aberration (LCA) between Eyes with Intraocular Lenses from Different Manufacturers.

Authors:  Masashi Nakajima; Takahiro Hiraoka; Toshiya Yamamoto; Seiu Takagi; Yoko Hirohara; Tetsuro Oshika; Toshifumi Mihashi
Journal:  PLoS One       Date:  2016-06-03       Impact factor: 3.240

3.  Through-Focus Energy Efficiency and Longitudinal Chromatic Aberration of Three Presbyopia-Correcting Intraocular Lenses.

Authors:  María S Millán; Fidel Vega
Journal:  Transl Vis Sci Technol       Date:  2020-11-09       Impact factor: 3.283

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.