| Literature DB >> 24257493 |
Tetsuyoshi Horiuchi1, Kiyoshi Ito, Kazuhiro Hongo, Masato Shibuya.
Abstract
Long titanium aneurysm clips have recently been released. In the present study, comparative study of long titanium and cobalt alloy clips was performed. Two kinds of Sugita long clips (straight clips of 25- and 35-mm blade length) made of titanium and cobalt alloys were tested by measuring the closing force, the anti-scissoring torque, and the maximum opening width. There were some differences between the two materials. In the 25-mm blade length clip, closing force and maximum opening width of titanium alloy clip were greater than those of cobalt alloy clip. By contrast, the anti-scissoring torque of 35-mm blade length titanium clip was stronger than that of the cobalt. The long titanium clips would have equivalent endurance to long cobalt clip and are safe for clinical use.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24257493 PMCID: PMC4533422 DOI: 10.2176/nmc.oa2013-0042
Source DB: PubMed Journal: Neurol Med Chir (Tokyo) ISSN: 0470-8105 Impact factor: 1.742
Fig. 1Photographs and schematic drawings show titanium (17-001-72 and 17-001-91) and cobalt long clips (07-940-72 and 07-940-91).
Characteristics of the aneurysm clips used in the present study
| Product name or material | Catalog number | Shape | Purpose | Blade length | Applied number |
|---|---|---|---|---|---|
| Titanium II | 17-001-72 | Straight | Permanent | 25 | 5 |
| Elgiloy | 07-940-72 | Straight | Permanent | 25 | 5 |
| Titanium II | 17-001-91 | Straight | Permanent | 35 | 5 |
| Elgiloy | 07-940-91 | Straight | Permanent | 35 | 5 |
Fig. 2Effect of distance from tip on Titanium II and Elgiloy closing forces. left: 25-mm blade clip (17-001-72 and 07-940-72) and right: 35-mm blade clip (17-001-92 and 07-940-92). Asterisk indicates significant differences between Titanium II and Elgiloy.
Fig. 3Effects of torque value during Titanium II and Elgiloy clip head rotations. A: 25-mm blade clip and B: 35-mm blade clip. Asterisk indicates significant differences between Titanium II and Elgiloy.
Fig. 4Photographs show the maximum opening width. upper column: 25-mm blade clip, and lower column: 35-mm blade clip.