Literature DB >> 24257381

Apparent dead space with the anesthetic conserving device, AnaConDa®: a clinical and laboratory investigation.

Louise W Sturesson1, Mikael Bodelsson, Anders Johansson, Björn Jonson, Gunnar Malmkvist.   

Abstract

BACKGROUND: The anesthetic conserving device (ACD) reduces consumption of volatile anesthetic drug by a conserving medium adsorbing exhaled drug during expiration and releasing it during inspiration. Elevated arterial CO2 tension (PaCO2) has been observed in patients using the ACD, despite tidal volume increase to compensate for larger apparatus dead space. In a test lung using room temperature dry gas, this was shown to be due to adsorption of CO2 in the ACD during expiration and release of CO2 during the following inspiration. The effect in the test lung was higher than in patients. We tested the hypothesis that a lesser dead space effect in patients is due to higher temperature and/or moisture attenuating rebreathing of CO2.
METHODS: The lungs of 6 postoperative cardiac surgery patients were ventilated using a conventional heat and moisture exchanger (HME) or an ACD. The ACD was studied with a test lung at varying temperatures and moistures. Infrared spectrometry was used to measure apparent dead space by the single-breath test for CO2 as well as rebreathing of CO2.
RESULTS: In patients, the median apparent dead space was 136 mL (95% confidence interval [CI,] 120-167) larger using the ACD compared with an HME (after correction for difference in internal volume 100 and 50 mL, respectively). Median rebreathing of CO2 using the ACD was 53% (range 48-58) of exhaled CO2 compared with 29% (range 27-32) with an HME. The median difference in CO2 rebreathing was 23% (95% CI, 18-27). In the test lung apparent dead space using ACD was unaffected by body temperature but decreased from 360 to 260 mL when moisture was added. This decreased rebreathing of CO2 from 62% to 48%.
CONCLUSIONS: The use of an ACD increases apparent dead space to a greater extent than can be explained by its internal volume. This is caused by adsorption of CO2 in the ACD during expiration and release of CO2 during inspiration. Rebreathing of CO2 was attenuated by moisture. The dead space effect of the ACD could be clinically relevant in acute respiratory distress syndrome and other diseases associated with ventilation difficulties, but investigations with larger sample sizes would be needed to determine the clinical importance.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24257381     DOI: 10.1213/ANE.0b013e3182a7778e

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  6 in total

1.  Uncontrolled delivery of liquid volatile anaesthetic when using the anaesthetic conserving device.

Authors:  Igor Karnjuš; Dušan Mekiš; Miljenko Križmarić
Journal:  J Clin Monit Comput       Date:  2017-05-31       Impact factor: 2.502

Review 2.  [New technical developments for inhaled sedation].

Authors:  A Meiser; H Bomberg; T Volk; H V Groesdonk
Journal:  Anaesthesist       Date:  2017-04       Impact factor: 1.041

3.  Halving the volume of AnaConDa: initial clinical experience with a new small-volume anaesthetic reflector in critically ill patients-a quality improvement project.

Authors:  Hagen Bomberg; Franziska Meiser; Sarah Zimmer; Martin Bellgardt; Thomas Volk; Daniel I Sessler; Heinrich V Groesdonk; Andreas Meiser
Journal:  J Clin Monit Comput       Date:  2018-04-26       Impact factor: 2.502

4.  Volumetric and reflective device dead space of anaesthetic reflectors under different conditions.

Authors:  Hagen Bomberg; Max Veddeler; Thomas Volk; Heinrich V Groesdonk; Andreas Meiser
Journal:  J Clin Monit Comput       Date:  2018-01-27       Impact factor: 2.502

5.  Comparison of the use of AnaConDa® versus AnaConDa-S® during the post-operative period of cardiac surgery under standard conditions of practice.

Authors:  Jose Miguel Marcos-Vidal; María Merino; Rafael González; Cristina García; Saleta Rey; Irene Pérez
Journal:  J Clin Monit Comput       Date:  2019-02-19       Impact factor: 2.502

Review 6.  A technical review of the history, development and performance of the anaesthetic conserving device "AnaConDa" for delivering volatile anaesthetic in intensive and post-operative critical care.

Authors:  Ron Farrell; Glen Oomen; Pauric Carey
Journal:  J Clin Monit Comput       Date:  2018-01-31       Impact factor: 2.502

  6 in total