| Literature DB >> 24256323 |
R Paseiro-Cerrato1, G O Noonan, T H Begley.
Abstract
Primary aromatic amines (PAAs) are a group of substances with undesirable health effects, that are used in a variety of commercial products. Several recent studies, using a number of screening and confirmatory methods, have reported the migration of PAAs from some kitchen utensils into acetic acid 3% (w/v). Many of these methods require significant sample preparation, therefore the aim of this work was to determine if direct analysis in real time mass spectrometry (DART-MS) could be utilised as a rapid screening tool for the determination of PAAs in kitchen utensils. DART-MS results from direct analysis of the utensil have been compared with results of PAA migration by ultra high performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method. The UPLC-MS/MS method had excellent linearity, appropriate sensitivity (LOD ≤ 1.5 µg L(-1); LOQ ≤ 4.5 µg L(-1)), repeatability from 2.4 to 13.2% and acceptable recoveries. DART-MS results were in good agreement with UPLC-MS/MS data, with 100% of non-compliant (PAA positive) samples successfully identified by DART-MS.Entities:
Mesh:
Substances:
Year: 2014 PMID: 24256323 DOI: 10.1080/19440049.2013.867366
Source DB: PubMed Journal: Food Addit Contam Part A Chem Anal Control Expo Risk Assess ISSN: 1944-0057