| Literature DB >> 24239967 |
Maarten P J Kuenen, Ingeborg H F Herold, Hendrikus H M Korsten, Jean J M C H de la Rosette, Hessel Wijkstra, Massimo Mischi.
Abstract
Indicator-dilution methods are widely used by many medical imaging techniques and by dye-, lithium-, and thermodilution measurements. The measured indicator dilution curves are typically fitted by a mathematical model to estimate the hemodynamic parameters of interest. This paper presents a new maximum-likelihood algorithm for parameter estimation, where indicator dilution curves are considered as the histogram of underlying transit-time distribution. Apart from a general description of the algorithm, semianalytical solutions are provided for three well-known indicator dilution models. An adaptation of the algorithm is also introduced to cope with indicator recirculation. In simulations as well as in experimental data obtained by dynamic contrast-enhanced ultrasound imaging, the proposed algorithm shows a superior parameter estimation accuracy over nonlinear least-squares regression. The feasibility of the algorithm for use in vivo is evaluated using dynamic contrast-enhanced ultrasound recordings obtained with the purpose of prostate cancer detection. The proposed algorithm shows an improved ability (increase in receiver-operating characteristic curve area of up to 0.13) with respect to existing methods to differentiate between healthy tissue and cancer.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24239967 DOI: 10.1109/TBME.2013.2290375
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538