Inductively coupled plasma mass spectrometry with an enlarged sampling orifice and offset Ion Lens. I. Ion trajectories and detector performance.

K Hu1, P S Clemons, R S Houk.   

Abstract

A new inductively coupled plasma mass spectrometer (ICP-MS) with four stages of differential pumping is described. The relatively large sampling orifice (1.31-mm dia.) improves signals for metal ions and resists plugging from deposited solids. A new ion lens is described that deflects ions off center and then back on center into the differential pumping orifice; there is no photon stop in the center of the beam. Calculations of ion trajectories using SIMIOn show that only those ions that leave the skimmer on center are transmitted, whereas most other lenses used in ICP-MS transmit only ions that leave the skimmer off axis. The performance of a Channeltron electron multiplier is compared to that of a Daly detector. Both detectors yield similar sensitivities of ≈ 10(6) counts s(-1) per ppm and detection limits of ≈ 1 pptr. The background with a Channeltron electron multiplier is only 0.4 counts s(-1) and is only slightly higher than the dark current count rate. Presumably the offset ion lens used in the present work efficiently screens the detector from photons emitted by the plasma. The background with the Daly detector is 4 counts s(-1), which represents a substantial improvement over the background obtained in previous use of the Daly detector with JCP-MS.

Entities:  

Year:  1993        PMID: 24234740     DOI: 10.1016/1044-0305(93)85038-Y

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  4 in total

1.  Effects of target gas in collision-induced dissociation using a double quadrupole mass spectrometer and radiofrequency.

Authors:  G Discharge; Y Mei; D C Duckworth; P R Cable; R K Marcus
Journal:  J Am Soc Mass Spectrom       Date:  1994-09       Impact factor: 3.109

2.  Determination of carbon isotope ratios in amino acids, proteins, and oligosaccharides by inductively coupled plasma-mass spectrometry.

Authors:  Elise T Luong; R S Houk
Journal:  J Am Soc Mass Spectrom       Date:  2003-04       Impact factor: 3.109

3.  Inductively coupled plasma mass spectrometry with an electrically floating sampling interface.

Authors:  K Hu; R S Hauk
Journal:  J Am Soc Mass Spectrom       Date:  1993-09       Impact factor: 3.109

Review 4.  Facets of ICP-MS and their potential in the medical sciences-Part 1: fundamentals, stand-alone and hyphenated techniques.

Authors:  David Clases; Raquel Gonzalez de Vega
Journal:  Anal Bioanal Chem       Date:  2022-08-27       Impact factor: 4.478

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.