| Literature DB >> 24234503 |
Abstract
A new event for the Fourier transform mass spectrometry (FTMS) sequence is developed and demonstrated. During this event, called a radiofrequency (RF)-only mode event, the typical passive cubic trap of a Fourier transform mass spectrometer is made to operate as an active quadrupole ion trap. The transition between active and passive modes is developed so that ion loss as a consequence of the transition can be held to 15% or less. The adduct of the ion-molecule reaction of the 1,3-butadiene radical cation and methyl vinyl ether was detected during the Rf-only-mode event at a helium pressure of ∼1×10(-3) torr even though this adduct is not detectable under standard FTMS operating conditions.Entities:
Year: 1992 PMID: 24234503 DOI: 10.1016/1044-0305(92)85037-K
Source DB: PubMed Journal: J Am Soc Mass Spectrom ISSN: 1044-0305 Impact factor: 3.109