K R Mohan1, M G Bartlett, K L Busch, A E Schoen, N Gore.
Abstract
A protocol for establishing standard instrument conditions for measurement of product ion MS/MS spectra from parent ions produced by electron ionization is presented. Within this protocol, the ion at m/z 231 (C5F9 (+)) from perfluorokerosene or perfluorotributylamine is selected as the parent ion and subjected to collision-induced dissociation. The relative intensities of product ions at m/z 69, 131, and 181 are monitored as a function of collision energy while keeping the target gas pressure constant within the range of 10(-4)-10(-6) torr (measured), or a beam attenuation of approximately 30-70%. The collision energy at which the ion intensities for product ions at m/z 69 and 181 are equal is defined as the calibration point at that collision gas pressure; the intensity of the ion at m/z 131 is very close to this value as well. Electron ionization MS/MS spectra taken at the calibration point using two different multiquadrupole instruments show good reproducibility for several test compounds. The high degree of similarity may aid in the establishment of a MS/MS spectral library.Entities:
Year: 1994 PMID: 24222626 DOI: 10.1016/1044-0305(94)90007-8