Literature DB >> 24218642

Low-level gamma and neutron monitoring based on use of proportional counter filled with 3He in polythene moderator: study of the responses to gamma and neutrons.

S Pszona1, A Bantsar2, P Tulik2, K Wincel2, B Zaręba2.   

Abstract

It has been shown that a proportional counter filled with (3)He placed centrally inside a polythene sphere opens a new possibility for measuring gamma photons and neutrons in the separate pulse-height windows. The responses to gamma and neutrons (in terms of ambient dose equivalent) of the detector assembly consisting of 203-mm polythene sphere with centrally positioned 40-mm diameter (3)He proportional counter have been studied. The response to secondary gammas from capture process in hydrogen has also been studied. The rather preliminary studies indicate that the proposed measuring system has very promising features as an ambient dose equivalent device for mixed gamma-neutron fields.
© The Author 2013. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24218642     DOI: 10.1093/rpd/nct274

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  Characteristics of Protons Exiting from a Polyethylene Converter Irradiated by Neutrons with Energies between 1 keV and 10 MeV.

Authors:  D Nikezic; Mehrdad Shahmohammadi Beni; D Krstic; K N Yu
Journal:  PLoS One       Date:  2016-06-30       Impact factor: 3.240

  1 in total