Literature DB >> 24214861

Alternating current corona discharge/atmospheric pressure chemical ionization for mass spectrometry.

Ahsan Habib1, Dilshadbek Usmanov, Satoshi Ninomiya, Lee Chuin Chen, Kenzo Hiraoka.   

Abstract

RATIONALE: Although alternating current (ac) corona discharge has been widely used in the fields of material science and technology, no reports have been published on its application to an atmospheric pressure chemical ionization (APCI) ion source. In this work, ac corona discharge for an APCI ion source has been examined for the first time.
METHODS: The ambient atmospheric pressure ac corona discharge (15 kHz, 2.6 kVptp ) was generated by using a stainless steel acupuncture needle. The generated ions were measured using an ion trap mass spectrometer. A comparative study on ac and direct current (dc) corona APCI ion sources was carried out using triacetone triperoxide and trinitrotoluene as test samples.
RESULTS: The ac corona discharge gave ion signals as strong as dc corona discharge for both positive and negative ion modes. In addition, softer ionization was obtained with ac corona discharge than with dc corona discharge. The erosion of the needle tip induced by ac corona was less than that obtained with positive mode dc corona.
CONCLUSIONS: A good 'yardstick' for assessing ac corona is that it can be used for both positive and negative ion modes without changing the polarity of the high-voltage power supply. Thus, ac corona can be an alternative to conventional dc corona for APCI ion sources.
Copyright © 2013 John Wiley & Sons, Ltd.

Entities:  

Year:  2013        PMID: 24214861     DOI: 10.1002/rcm.6744

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  5 in total

1.  Gas Chromatography-Tandem Mass Spectrometry of Lignin Pyrolyzates with Dopant-Assisted Atmospheric Pressure Chemical Ionization and Molecular Structure Search with CSI:FingerID.

Authors:  Evan A Larson; Carolyn P Hutchinson; Young Jin Lee
Journal:  J Am Soc Mass Spectrom       Date:  2018-06-12       Impact factor: 3.109

2.  Unexpected Reduction of Iminoquinone and Quinone Derivatives in Positive Electrospray Ionization Mass Spectrometry and Possible Mechanism Exploration.

Authors:  Jiying Pei; Cheng-Chih Hsu; Ruijie Zhang; Yinghui Wang; Kefu Yu; Guangming Huang
Journal:  J Am Soc Mass Spectrom       Date:  2017-08-07       Impact factor: 3.109

3.  Enhanced aerodynamic reach of vapor and aerosol sampling for real-time mass spectrometric detection using Venturi-assisted entrainment and ionization.

Authors:  Thomas P Forbes; Matthew Staymates
Journal:  Anal Chim Acta       Date:  2017-01-02       Impact factor: 6.558

4.  Portable Digital Linear Ion Trap Mass Spectrometer Based on Separate-Region Corona Discharge Ionization Source for On-Site Rapid Detection of Illegal Drugs.

Authors:  Lingfeng Li; Tianyi Zhang; Deting Wang; Yunjing Zhang; Xingli He; Xiaozhi Wang; Peng Li
Journal:  Molecules       Date:  2022-05-30       Impact factor: 4.927

5.  Corona Discharge and Field Electron Emission in Ambient Air Using a Sharp Metal Needle: Formation and Reactivity of CO3 -• and O2 -•.

Authors:  Kenzo Hiraoka; Stephanie Rankin-Turner; Satoshi Ninomiya; Haruo Shimada; Kazumasa Kinoshita; Shinichi Yamabe
Journal:  Mass Spectrom (Tokyo)       Date:  2021-12-25
  5 in total