How to measure the kinetic energy distribution of the precursor ion main beam in mass-analyzed ion kinetic energy spectrometry experiments.

F H Strobel1, J Adams.   

Abstract

Scans of the electrostatic analyzer (ESA) across the precursor ion beam in reverse-geometry (BE) mass spectrometers that are operated under double-focusing conditions do not measure the "energy resolution of the main beam": They only measure double-focusing resolution. The only way that ESA scans can measure the kinetic energy distribution of the main beam is to operate the instrument so that angular (directional) focusing is not achieved. Thus, the mass spectrometer is no longer double-focusing. Under double-focusing conditions, however, scans of the accelerating voltage while the magnetic field and ESA are held constant can be used to measure either the kinetic energy distribution of the main beam that enters the magnet or the energy-resolving power of the instrument. Scans at a constant ratio of B(2)/E can be used similarly. The energy-resolving power of any ESA is defined by its dispersion and the widths of the energy-resolving object and image slits that immediately precede and follow the ESA, respectively. The use of BE, EB, and triple-sector instruments to measure energy-resolving power and the kinetic energy distribution of the precursor ion main beam is compared and discussed.

Year:  1995        PMID: 24214075     DOI: 10.1016/1044-0305(95)00570-6

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  2 in total

1.  Primary and secondary locations of charge sites in angiotensin II (M + 2H)2+ ions formed by electrospray ionization.

Authors:  M C Sullards; J A Reiter
Journal:  J Am Soc Mass Spectrom       Date:  2000-01       Impact factor: 3.109

2.  The importance of charge-separation reactions in tandem mass spectrometry of doubly protonated angiotensin II formed by electrospray ionization: Experimental considerations and structural implications.

Authors:  J Adams; F H Strobel; A Reiter; M C Sullards
Journal:  J Am Soc Mass Spectrom       Date:  1996-01       Impact factor: 3.109

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.