Internal pulsed valve sample introduction on a quadrupole ion trap mass spectrometer.

B I Coopersmith1, R A Yost.   

Abstract

A pulsed valve positioned just outside the ion trap electrodes (within the vacuum chamber) has been characterized. The observed gas pulse widths and the maximum ion intensities were found to decrease as the distance between the pulsed valve and the ion trap electrodes increased. An explanation is presented within. The pulsed valve was found to impart temporal separation in ion-molecule reactions by permitting the removal of interfering neutrals. Other factors that affect the degree of temporal separation also are presented.

Year:  1995        PMID: 24214042     DOI: 10.1016/1044-0305(95)00528-L

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  8 in total

1.  Pulsed gas introduction into quadrupole ion traps.

Authors:  W B Emary; R E Kaiser; H I Kenttämaa; R G Cooks
Journal:  J Am Soc Mass Spectrom       Date:  1990-07       Impact factor: 3.109

2.  Alternating RF/DC Isolations for Quantitation with Coeluting Internal Standards in Gas Chromatography/Ion Trap Mass Spectrometry.

Authors:  B L Kleintop; R A Yost; C R Abolin
Journal:  J Am Soc Mass Spectrom       Date:  1992-01       Impact factor: 3.109

3.  Fourier-transform electrospray instrumentation for tandem high-resolution mass spectrometry of large molecules.

Authors:  S C Beu; M W Senko; J P Quinn; F M Wampler; F W McLafferty
Journal:  J Am Soc Mass Spectrom       Date:  1993-07       Impact factor: 3.109

4.  Gas-phase ion-molecule reactions: a model for the determination of biologically reactive electrophilic contaminants in the environment.

Authors:  J A Freeman; J V Johnson; R A Yost; D W Kuehl
Journal:  Anal Chem       Date:  1994-06-01       Impact factor: 6.986

5.  Functional group-selective ion-molecule reactions of ethylene glycol and its monomethyl and dimethyl ethers.

Authors:  E S Eichmann; J S Brodbelt
Journal:  J Am Soc Mass Spectrom       Date:  1993-02       Impact factor: 3.109

6.  Selective detection of the tolyl cation among other [C7H 7] (+) isomers by ion/molecule reaction with dimethyl ether.

Authors:  T G Heath; J Allison; J T Watson
Journal:  J Am Soc Mass Spectrom       Date:  1991-08       Impact factor: 3.109

7.  Information on the location of carbon-carbon double bonds in C6-C 23 linear alkenes from carbon addition reactions in a quadrupole ion trap equipped with a pulsed sample-inlet system.

Authors:  J Einhorn; H I Kenttämaa; R G Cooks
Journal:  J Am Soc Mass Spectrom       Date:  1991-08       Impact factor: 3.109

8.  Estimation of mutagenic/carcinogenic potential of environmental contaminants by ion-molecule reactions and tandem mass spectrometry.

Authors:  J A Freeman; J V Johnson; M E Hail; R A Yost; D W Kuehl
Journal:  J Am Soc Mass Spectrom       Date:  1990-02       Impact factor: 3.109

  8 in total
  1 in total

1.  MS/MS of ions in a low pressure linear ion trap using a pulsed gas.

Authors:  Bruce A Collings; Matthew A Romaschin
Journal:  J Am Soc Mass Spectrom       Date:  2009-05-20       Impact factor: 3.109