| Literature DB >> 24211877 |
Abstract
A highly sensitive chemiluminescent immunoassay (CLIA) using a sensitive organic photodetector was developed to detect human cortisol, an important biomarker for stress-related diseases. The developed CLIA was performed onto gold-coated glass chips, on which anti-cortisol antibodies were immobilised and chemiluminescent horseradish peroxidase-luminol-peroxide reactions were generated. Using cortisol-spiked artificial saliva samples, the CLIA biosensor showed a linear range of detection between 0.1 ng/mL and 175 ng/mL and a detection limit of 80 pg/mL. The sensor response was highly specific to cortisol and did not vary significantly between assays. The results indicate the potential clinical application of the CLIA sensor. Furthermore, the simple layered structure of the organic photodetector may encourage the realisation of integrated optical biosensors for point-of-use measurement of salivary cortisol levels.Entities:
Keywords: Thin-film organic electronics; chemiluminescent immunoassay; optical biosensor; salivary cortisol
Mesh:
Substances:
Year: 2014 PMID: 24211877 DOI: 10.3233/BME-130778
Source DB: PubMed Journal: Biomed Mater Eng ISSN: 0959-2989 Impact factor: 1.300