A Rohlmann1, M Dreischarf2, T Zander2, F Graichen2, G Bergmann2.
Abstract
Walking is one of the most important activities in daily life, and walking exposes the spine to a high number of loading cycles. Little is known about the spinal loads during walking. Telemeterized spinal implants can provide data about their loading during different activities. The aim of this study was to measure the loads on a vertebral body replacement (VBR) during level and staircase walking and to determine the effects of walking speed and using walking aids. Telemeterized VBRs were implanted in five patients suffering from compression fractures of the L1 or L3 lumbar vertebral body. The implant allows measurements of three force and three moment components. The resultant force on the VBR was measured during level and staircase walking, when walking on a treadmill at different speeds, and when using a wheeled invalid walker or crutches. On average, the resultant force on the VBR for level walking was 171% of the value for standing. This force value increased to 265% of the standing force when ascending stairs and to 225% when descending stairs. Walking speed had a strong effect on the implant force. Using a walker during ambulation on level ground reduced the force on the implant to 62% of standing forces, whereas using two crutches had only a minor effect. Walking causes much higher forces on the VBR than standing. A strong force reduction can be achieved by using a walker.Entities:
Keywords: Load measurement; Spinal loads; Telemetry; Vertebral body replacement; Walking
Mesh:
Year: 2013 PMID: 24211089 DOI: 10.1016/j.gaitpost.2013.10.010
Source DB: PubMed Journal: Gait Posture ISSN: 0966-6362 Impact factor: 2.840