Early postoperative failure of a new intramedullary fixation device for midshaft clavicle fractures.

David J Wilson, DeWayne L Weaver, Todd P Balog, Edward D Arrington.   

Abstract

The Sonoma CRx device (Sonoma Orthopedic Products, Santa Rosa, California) is a recently introduced intramedullary device with a flexible shaft that becomes rigid once actuated to allow deployment within the sigmoidal contour of the clavicular shaft. Medial intramedullary cortical purchase is obtained by grippers and lateral purchase through a locking bicortical buttressing screw. This article describes 2 cases of early hardware failure using this device. In both cases, early postoperative radiographs demonstrate adequate initial fracture reduction and implant position. Both patients sustained repeat injuries, one under low physiologic load and the other after returning to mixed martial arts 4 months postoperatively. Implant failure was noted after reinjury in both cases. Complete healing and full return to function was documented for both patients at 2 years. Proper patient selection and counseling regarding the limitations of this intramedullary fixation device are important. Biomechanical comparison of this implant to plate fixation under physiologic loads of combined axial compression and torsion may shed light on differences in fixation stability. Copyright 2013, SLACK Incorporated.

Entities:  

Mesh:

Year:  2013        PMID: 24200452     DOI: 10.3928/01477447-20131021-31

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  3 in total

1.  Minimally invasive implantation of a novel flexible intramedullary nail in patients with displaced midshaft clavicle fractures.

Authors:  M Calbiyik; S Zehir; D Ipek
Journal:  Eur J Trauma Emerg Surg       Date:  2015-08-29       Impact factor: 3.693

2.  Minimally invasive fixation of midclavicular fractures with threaded elastic intramedullary nails.

Authors:  Hongbin Bi; Yongqing Wang; Qingguang Xiong; Yi Li; Zhihui Zhao; Zhiqiang Yang
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-04-18

3.  Prospective randomized study comparing results of fixation for clavicular shaft fractures with intramedullary nail or locking compression plate.

Authors:  Murat Calbiyik; Deniz Ipek; Mehmet Taskoparan
Journal:  Int Orthop       Date:  2016-05-02       Impact factor: 3.075

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.