Feasibility of insertion of a microcatheter through a Y-stent in coil embolization of cerebral aneurysms and its detailed geometry by micro-computed tomography.

Kenichi Kono1, Tomoaki Terada.   

Abstract

BACKGROUND: In Y-stent-assisted coil embolization for cerebral aneurysms, open or closed cell stents are used. Different microcatheters for coil insertion are available. We investigated which microcatheter could be navigated into an aneurysm through a Y-stent with different stents.
METHODS: Double Neuroform open-cell stents or double Enterprise closed-cell stents were deployed in Y-configuration in a silicon model of a bifurcation aneurysm. Two endovascular neurosurgeons independently tried to navigate an SL-10 microcatheter for 0.010" coils or a PX Slim microcatheter for 0.020" Penumbra coils into the aneurysm through the Y-stent. In addition, we measured lengths of stent pores of the Y-stents with double Enterprise stents deployed in the model by micro-computed tomography.
RESULTS: It was feasible to navigate an SL-10 microcatheter into the aneurysm through the Y-stent with Enterprise or Neuroform stents. Navigation of a PX Slim microcatheter was feasible in the Y-stents only with Neuroform stents. In the Y-stent with double Enterprise stents, the lengths of the second stent pores were significantly smaller than those of the first stent (0.41 ± 0.18 mm vs 0.69 ± 0.20 mm; P = 0.008). The SL-10 microcatheter was smaller than approximately 80% of the stent pores of the first stent and 30% of those of the second stent. The PX Slim microcatheter was smaller than 20% of the stent pores of the first stent and 0% of those of the second stent.
CONCLUSIONS: It was feasible to insert an SL-10 microcatheter into the aneurysm through Y-stents with Enterprise or Neuroform stents. Navigation of a PX Slim microcatheter for 0.020" Penumbra coils was feasible in Y-stents with Neuroform stents, but not with double Enterprise stents. The measurements of stent pores by micro-computed tomography supported this feasibility study. These results may be helpful to select appropriate stents and microcatheters in Y-stent-assisted coil embolization, especially in case of retreatments.

Entities:  

Mesh:

Year:  2013        PMID: 24190455     DOI: 10.1007/s00701-013-1925-4

Source DB:  PubMed          Journal:  Acta Neurochir (Wien)        ISSN: 0001-6268            Impact factor:   2.216


  4 in total

1.  Computational fluid dynamics of cerebral aneurysm coiling using high-resolution and high-energy synchrotron X-ray microtomography: comparison with the homogeneous porous medium approach.

Authors:  Michael R Levitt; Michael C Barbour; Sabine Rolland du Roscoat; Christian Geindreau; Venkat K Chivukula; Patrick M McGah; John D Nerva; Ryan P Morton; Louis J Kim; Alberto Aliseda
Journal:  J Neurointerv Surg       Date:  2016-07-12       Impact factor: 5.836

2.  Bioresorbable Vascular Scaffolds in Interventional Neuroradiology.

Authors:  Anil Arat; Ergun Daglioglu; Ilkay Akmangit; Ahmet Peker; Murat Arsava; Mehmet Akif Topcuoglu; Deniz Belen
Journal:  Clin Neuroradiol       Date:  2017-07-26       Impact factor: 3.649

3.  Novel Noncrossing Y-Stent Technique Using Tapered Proximal End of a Solitaire AB Stent for Coil Embolization of Wide-Neck Bifurcation Aneurysms.

Authors:  Hyon-Jo Kwon; Jeong-Wook Lim; Hyoung Soo Byoun; Hyeon-Song Koh
Journal:  J Korean Neurosurg Soc       Date:  2020-07-09

4.  Hemodynamic effects of stent struts versus straightening of vessels in stent-assisted coil embolization for sidewall cerebral aneurysms.

Authors:  Kenichi Kono; Aki Shintani; Tomoaki Terada
Journal:  PLoS One       Date:  2014-09-23       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.