| Literature DB >> 24182151 |
Sergey A Telenkov1, Rudolf Alwi, Andreas Mandelis.
Abstract
Photoacoustic (PA) imaging of biological tissues using laser diodes instead of conventional Q-switched pulsed systems provides an attractive alternative for biomedical applications. However, the relatively low energy of laser diodes operating in the pulsed regime, results in generation of very weak acoustic waves, and low signal-to-noise ratio (SNR) of the detected signals. This problem can be addressed if optical excitation is modulated using custom waveforms and correlation processing is employed to increase SNR through signal compression. This work investigates the effect of the parameters of the modulation waveform on the resulting correlation signal and offers a practical means for optimizing PA signal detection. The advantage of coherent signal averaging is demonstrated using theoretical analysis and a numerical model of PA generation. It was shown that an additional 5-10 dB of SNR can be gained through waveform engineering by adjusting the parameters and profile of optical modulation waveforms.Entities:
Mesh:
Year: 2013 PMID: 24182151 DOI: 10.1063/1.4825034
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523