Literature DB >> 24177042

The influence of SPECT reconstruction algorithms on image quality and diagnostic accuracy in phantom measurements and 99mTc-sestamibi parathyroid scintigraphy.

Ralph A van Hoorn1, Dennis Vriens, Jan-Willem A Postema, Anne I J Arens, Andreas Pfestroff, Wim J G Oyen, Martin Gotthardt.   

Abstract

OBJECTIVE: The aim of the study was to determine the influence and clinical consequences of different tomographic reconstruction algorithms on the image quality and diagnostic accuracy of low count statistics single photon emission computed tomography (SPECT).
METHODS: Phantom scans were used to assess the image quality (percentage of recovered contrast, relative noise level, and spatial resolution). The diagnostic accuracy of parathyroid SPECT performed in 60 patients with primary hyperparathyroidism was assessed. Phantom and patient SPECT scans were reconstructed by two different algorithms (ReSPECT and HOSEM). Two blinded nuclear medicine physicians interpreted the patient scans in random order, without knowledge of any clinical data. Subjective image quality and certainty of diagnosis were scored.
RESULTS: Significantly lower relative noise levels (0.08 vs. 0.13, respectively; P=0.042) and higher image spatial resolution (6.6 vs. 17.1 mm full-width at half-maximum in the transaxial plane, respectively) were found in the phantom studies using ReSPECT as compared with HOSEM. In the clinical scans, the mean target-to-background ratio was higher for ReSPECT (4.53 vs. 2.97; P<0.001). There was no significant difference in region-based sensitivity (73 vs. 68%; P=0.538) or specificity (93 vs. 92%; P=0.815) between ReSPECT and HOSEM. Subjective image quality as well as certainty of interpretation was higher for ReSPECT (P<0.001).
CONCLUSION: There is a significant difference in image quality between commercially available algorithms for tomographic reconstruction of SPECT scans, which appears to have an impact on diagnostic accuracy and interpreter certainty. Therefore, we believe that a systematic comparison of different SPECT reconstruction algorithms should be made to ensure reproducible high image quality in clinical SPECT.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24177042     DOI: 10.1097/MNM.0000000000000020

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  3 in total

1.  99mTc-Sestamibi/123I Subtraction SPECT/CT in Parathyroid Scintigraphy: Is Additional Pinhole Imaging Useful?

Authors:  Virpi Tunninen; Pekka Varjo; Tomi Kauppinen; Anu Holm; Hannu Eskola; Marko Seppänen
Journal:  Int J Mol Imaging       Date:  2017-10-18

2.  Enhanced Specific Activity by Multichelation of Exendin-3 Leads To Improved Image Quality and In Vivo Beta Cell Imaging.

Authors:  Lieke Joosten; Maarten Brom; Hanneke Peeters; Sandra Heskamp; Martin Béhé; Otto Boerman; Martin Gotthardt
Journal:  Mol Pharm       Date:  2017-12-21       Impact factor: 4.939

3.  Influence of single photon emission computed tomography (SPECT) reconstruction algorithm on diagnostic accuracy of parathyroid scintigraphy: Comparison of iterative reconstruction with filtered backprojection.

Authors:  Gonca Kara Gedik; Oktay Sari
Journal:  Indian J Med Res       Date:  2017-04       Impact factor: 2.375

  3 in total