Analysis of the misfit of dental implant-supported prostheses made with three manufacturing processes.

Marc Fernández1, Luis Delgado1, Meritxell Molmeneu1, David García2, Daniel Rodríguez3.   

Abstract

STATEMENT OF PROBLEM: The microgap between implant components has been associated with complications such as screw loosening or adverse biologic responses.
PURPOSE: The purpose of this study was to quantify the microroughness of the mating surfaces of implant components manufactured with different processes, to quantify the microgap between implant components, and to determine whether a correlation exists between microroughness and the microgap.
MATERIAL AND METHODS: Nine dental implants with a standard external connection were paired with 3 milled, 3 cast, and 3 sintered compatible cobalt-chromium alloy abutments. The abutment surface was examined, and the roughness parameter Sz was measured by using a white-light interferometric microscope at ×10 to ×100 magnification. The abutment surface and the microgap of the implant-abutment connection were observed with scanning electron microscopy, and the microgap width was quantified from micrographs made of each implant-abutment pair. The mean and standard deviation of roughness and microgap were evaluated. A 1-way ANOVA (α=.05) was used to assess the influence of the manufacturing process on roughness and microgap. The Pearson correlation was used to check dependence between roughness and microgap.
RESULTS: The milled abutments possessed a connection geometry with defined edges and a mean roughness of 29 μm, sintered abutments showed a blurred but functional connection with a roughness of 115 μm, and cast abutments showed a connection with a loss of axial symmetry and a roughness of 98 μm. A strong correlation was found between the roughness values on the mating surfaces and the microgap width.
CONCLUSIONS: The milled components were smoother than the cast or sintered components. A correlation was found between surface roughness and microgap width.
Copyright © 2014 Editorial Council for the Journal of Prosthetic Dentistry. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24176182     DOI: 10.1016/j.prosdent.2013.09.006

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  4 in total

1.  Effect of Fabrication Technique on the Microgap of CAD/CAM Cobalt-Chrome and Zirconia Abutments on a Conical Connection Implant: An In Vitro Study.

Authors:  Pedro Molinero-Mourelle; Rocio Cascos-Sanchez; Burak Yilmaz; Walter Yu Hang Lam; Edmond Ho Nang Pow; Jaime Del Río Highsmith; Miguel Gómez-Polo
Journal:  Materials (Basel)       Date:  2021-04-30       Impact factor: 3.623

Review 2.  Abutment misfit in implant-supported prostheses manufactured by casting technique: An integrative review.

Authors:  Lorena M S Pereira; Mariane B Sordi; Ricardo S Magini; Antônio R Calazans Duarte; Júlio C M Souza
Journal:  Eur J Dent       Date:  2017 Oct-Dec

Review 3.  Trends in computer-aided manufacturing in prosthodontics: a review of the available streams.

Authors:  Jaafar Abduo; Karl Lyons; Mohammed Bennamoun
Journal:  Int J Dent       Date:  2014-04-08

4.  Role of span length in the adaptation of implant-supported cobalt chromium frameworks fabricated by three techniques.

Authors:  Ying Zhou; Yong Li; Xiao Ma; Yiqing Huang; Jiawei Wang
Journal:  J Adv Prosthodont       Date:  2017-04-06       Impact factor: 1.904