Optical coherence tomography evaluation of late strut coverage patterns between first-generation drug-eluting stents and everolimus-eluting stent.

Francisco Javier Toledano Delgado1, Manuel Pan Alvarez-Ossorio, José Suárez de Lezo Cruz-Conde, Francisco Mazuelos Bellido, Miguel Ángel Romero Moreno, Alfonso Medina Fernández-Aceytuno, Javier Suárez de Lezo Herrerosde Tejada, Soledad Ojeda Pineda, José María Segura Saint-Gerons, Djordje Pavlovic.   

Abstract

OBJECTIVES: To compare strut coverage patterns between everolimus-eluting stent (EES) and first-generation drug-eluting stents (DES) at more than 12 months after successful implantation, using optical coherence tomography (OCT).
BACKGROUND: No sufficient OCT data has been reported comparing late strut coverage patterns between EES and first-generation DES. The favorable late results after EES implantation could be related to lower rates of uncovered and malapposed struts.
METHODS: A total of 66 DES (21 EES, 23 SES, and 22 PES) that were implanted at least 1 year in advance in 40 patients and met good late angiographic results were evaluated by OCT. The percentage of uncovered and malapposed struts, calculated as the ratio of uncovered or malapposed struts to total struts in all cross-sectional images per stent, was compared among the three groups.
RESULTS: A total of 35,061 struts were analyzed: 11,967 from EES, 11,855 from SES, and 11m239 from PES. The average tissue coverage thickness of the struts per stent was greater in EES than in SES and PES (109 ± 40 µm vs. 72 ± 27 µm and 83 ± 26 µm, respectively; P = 0.001). The percentage of uncovered struts (1.9 ± 4.1% in EES vs. 11.6 ± 12.7% in SES, P = 0.01 and vs. 7.1 ± 5.2% in PES, P < 0.001) and malapposed struts (0.1 ± 0.3% in EES vs. 1.8 ± 3.5% in SES, P = 0.01 and vs. 3.5 ± 5.1% in PES, P = 0.02) was much lower in EES than in first-generation DES, with no significant differences between SES and PES.
CONCLUSIONS: Late strut coverage patterns are not similar between EES and first-generation DES. EES showed a lower percentage of uncovered and malapposed struts.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  drug delivery; malapposed struts; neointimal coverage; optical coherence tomography; stent structure; uncovered struts

Mesh:

Substances:

Year:  2013        PMID: 24174291     DOI: 10.1002/ccd.25235

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

Review 1.  Stenting of coronary bifurcation lesions: a literature and technical review.

Authors:  Joo Myung Lee; Kyung Woo Park; Bon-Kwon Koo; Hyo-Soo Kim
Journal:  Curr Cardiol Rep       Date:  2015-06       Impact factor: 2.931

2.  Quantitative analysis of the side-branch orifice after bifurcation stenting using en-face processing of OCT images: a comparison between Xience V and Resolute Integrity stents.

Authors:  Yoshiyasu Minami; Zhao Wang; Aaron D Aguirre; Stephen Lee; Shiro Uemura; Tsunenari Soeda; Rocco Vergallo; Owen C Raffel; Peter Barlis; Tomonori Itoh; Hang Lee; James Fujimoto; Ik-Kyung Jang
Journal:  Coron Artery Dis       Date:  2016-01       Impact factor: 1.439

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.