A comparative study of intraocular pressure measurement by three tonometers in normal subjects.

S K Prabhakar, B S Mahesh, M Shanthamallappa.   

Abstract

INTRODUCTION: Contact Goldmann applanation tonometry (GAT) is gold standard for measuring intraocular pressure; however its routine use is limited by its non-portability and the need for a Slit Lamp Microscope. The Portable Perkins tonometer is also considered gold standard because it is based on the same principles as the GAT. The iCare is a newly introduced, portable, non-contact tonometer (NCT) that measures intraocular pressure (IOP) using a thin metallic probe.
OBJECTIVE: To evaluate reliability and accuracy of IOP measurements using iCare and Keeler Pulsair tonometers against Perkins tonometer. SUBJECTS AND METHODS: A comparative, randomized, prospective clinical study conducted on 166 eyes of 83 (n=83) subjects in the age group 14 to 71 years. The pressures were first recorded by iCare and Pulsair and then by Perkins. The SPSS 11.00 version was used for analysis.
RESULTS: Mean pressures and standard deviation (+/-SD) for iCare, Pulsair and Perkins were 14.62(+/- 2.47), 14.53(+/-3.36) and 13.06(+/-2.69) and the Standard Error of the Mean (SEM) was 0.27, 0.36 and 0.30 respectively. There was a good correlation between iCare and Perkins with statistically significant difference (r=0.610, p less than 0.05). Regression analysis was performed. Using the Bland-Altman analysis 95% Limits of Agreement (LoA) for iCare and Pulsair were determined as -6.1 to 2.9 and -4.5 to 7.5 respectively.
CONCLUSION: Although both tonometers overestimated the Perkins values, Pulsair showed a better agreement with Perkins tonometer than iCare tonometer. © NEPjOPH.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24172555     DOI: 10.3126/nepjoph.v5i2.8729

Source DB:  PubMed          Journal:  Nepal J Ophthalmol        ISSN: 2072-6805


  5 in total

1.  Application of Cornelian Cherry Iridoid-Polyphenolic Fraction and Loganic Acid to Reduce Intraocular Pressure.

Authors:  Dorota Szumny; Tomasz Sozański; Alicja Z Kucharska; Wojciech Dziewiszek; Narcyz Piórecki; Jan Magdalan; Ewa Chlebda-Sieragowska; Robert Kupczynski; Adam Szeląg; Antoni Szumny
Journal:  Evid Based Complement Alternat Med       Date:  2015-06-01       Impact factor: 2.629

2.  Goldmann Applanation Tonometry: Using a Red-Free Filter for Mass Screening.

Authors:  Shahram Bamdad; Yalda Vatan
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2017

Review 3.  Portable hardware & software technologies for addressing ophthalmic health disparities: A systematic review.

Authors:  Margarita Labkovich; Megan Paul; Eliott Kim; Randal A Serafini; Shreyas Lakhtakia; Aly A Valliani; Andrew J Warburton; Aashay Patel; Davis Zhou; Bonnie Sklar; James Chelnis; Ebrahim Elahi
Journal:  Digit Health       Date:  2022-05-06

4.  Association of intraocular pressure and postoperative nausea and vomiting after microvascular decompression - a prospective cohort study.

Authors:  Yuantao Hou; Hansheng Liang; Cungang Fan; Ruen Liu; Yi Feng
Journal:  BMC Anesthesiol       Date:  2022-04-30       Impact factor: 2.217

5.  Association of Diabetes With Central Corneal Thickness Among a Multiethnic Asian Population.

Authors:  Xiao-Yang Luo; Wei Dai; Miao-Li Chee; Yijin Tao; Jacqueline Chua; Nicholas Y Q Tan; Yih-Chung Tham; Tin Aung; Tien Yin Wong; Ching-Yu Cheng
Journal:  JAMA Netw Open       Date:  2019-01-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.