Does the precision of dual-energy X-ray absorptiometry for bone mineral density differ by sex?

Diane Krueger1, Nellie Vallarta-Ast2, Jessie Libber2, Ronald Gangnon3, Neil Binkley2.   

Abstract

Given larger bone size in men, bone mineral density (BMD) precision might differ between sexes. This study compared dual-energy X-ray absorptiometry BMD precision of 3 International Society for Clinical Densitometry-certified technologists in older men and women. Each technologist scanned a cohort of 30 men and 30 women (total n = 180) by using a Lunar iDXA densitometer (GE Healthcare, Madison, WI). Each volunteer had 2 lumbar spine and bilateral hip scans with repositioning between examinations. BMD least significant change was calculated. Age and body mass index did not differ between men and women. Mean height and weight were greater in men, 174.6 cm ± 6.9 and 81.6 kg ± 11.1 respectively, (p < 0.0001) than in women, 161.5 cm ± 5.9/69.1 kg ± 14.2, respectively. Bone area was greater in men (p < 0.0001) at all sites. BMD least significant change was statistically better (p < 0.05) in women at the mean total femur (0.014 vs 0.018 g/cm(2)) and left femoral neck (0.025 vs 0.038 g/cm(2)), but not different at either total femur, the right femoral neck, or lumbar spine (all p > 0.05). In conclusion, statistically significant male/female differences in BMD precision were observed at the mean total femur and left femoral neck. Given the small magnitude of difference in g/cm(2) and inconsistent pattern, that is, no right femoral neck difference, there is virtually no clinical difference in BMD precision between sexes. These data do not support a need for sex-specific precision analyses.
Copyright © 2014 The International Society for Clinical Densitometry. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bone density; DXA; precision; sex

Mesh:

Year:  2013        PMID: 24169083     DOI: 10.1016/j.jocd.2013.07.003

Source DB:  PubMed          Journal:  J Clin Densitom        ISSN: 1094-6950            Impact factor:   2.617


  1 in total

1.  Dual-Energy X-ray Absorptiometry Image Resolution and TBS Precision.

Authors:  Robert D Blank
Journal:  J Clin Densitom       Date:  2015 Apr-Jun       Impact factor: 2.617

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.