Literature DB >> 24167807

Microsurgical extraction of a malfunctioned pipeline embolization device following complete deployment.

Dale Ding1, Kenneth C Liu.   

Abstract

The Pipeline Embolization Device (PED) is an effective treatment approach for complex intracranial aneurysms. Intraprocedural complications during PED deployment are seldom reported. We report a rare complication of a PED malfunction identified immediately following complete deployment during endovascular treatment of a giant middle cerebral artery (MCA) bifurcation aneurysm. After multiple failed attempts at endovascular retrieval of the malfunctioned PED, the patient was taken for microsurgical extraction due to accumulation of thrombus on the proximal unopened portion of the stent and widespread distal dissemination of emboli. After removing the PED from the vessel lumen and resecting the giant aneurysm, we could not reanastamose the proximal MCA to the distal segment. The management of PED malfunction is poorly understood. While removal of an incompletely deployed PED may be undertaken with limited adverse effects, retrieval of a fully deployed PED is associated with a much higher risk of morbidity. Until larger case series of such complications better define the risks and benefits of endovascular or microsurgical retrieval of malfunctioned PEDs, the management of these rare intraprocedural complications will be based on the unique aspects of each individual case and the expertise of the treating neurointerventionalist.

Entities:  

Keywords:  Endovascular procedures; Intracranial aneurysm; Intraoperative complications; Stents; Stroke

Year:  2013        PMID: 24167807      PMCID: PMC3804665          DOI: 10.7461/jcen.2013.15.3.241

Source DB:  PubMed          Journal:  J Cerebrovasc Endovasc Neurosurg        ISSN: 2234-8565


  14 in total

1.  In situ removal of the pipeline embolization device: the 'corking' and 'pseudo-corking' techniques.

Authors:  Geoffrey P Colby; Juan F Gomez; Li-Mei Lin; Alexandra R Paul; Alexander L Coon
Journal:  J Neurointerv Surg       Date:  2012-02-23       Impact factor: 5.836

2.  Review of reported complications associated with the Pipeline Embolization Device.

Authors:  Kyle M Fargen; Gregory J Velat; Matthew F Lawson; J Mocco; Brian L Hoh
Journal:  World Neurosurg       Date:  2012-02-11       Impact factor: 2.104

3.  Endovascular rescue of a misshapen intracranial stent: report of two cases.

Authors:  Michael Chen; Lindsay Goldstein-Smith
Journal:  J Neurointerv Surg       Date:  2010-08-25       Impact factor: 5.836

4.  Unruptured intracranial aneurysms--risk of rupture and risks of surgical intervention.

Authors: 
Journal:  N Engl J Med       Date:  1998-12-10       Impact factor: 91.245

5.  Transvascular coil hooking procedure to retrieve an unraveled Guglielmi detachable coil: technical note.

Authors:  Christian Raftopoulos; Pierre Goffette; Rudolf F Billa; Pierre Mathurin
Journal:  Neurosurgery       Date:  2002-04       Impact factor: 4.654

Review 6.  Giant intracranial aneurysms: endovascular challenges.

Authors:  Richard J Parkinson; Christopher S Eddleman; H Hunt Batjer; Bernard R Bendok
Journal:  Neurosurgery       Date:  2006-11       Impact factor: 4.654

7.  The pipeline embolization device for the intracranial treatment of aneurysms trial.

Authors:  P K Nelson; P Lylyk; I Szikora; S G Wetzel; I Wanke; D Fiorella
Journal:  AJNR Am J Neuroradiol       Date:  2010-12-09       Impact factor: 3.825

8.  Immediate and midterm results following treatment of unruptured intracranial aneurysms with the pipeline embolization device.

Authors:  W McAuliffe; V Wycoco; H Rice; C Phatouros; T J Singh; J Wenderoth
Journal:  AJNR Am J Neuroradiol       Date:  2011-10-06       Impact factor: 3.825

9.  Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms.

Authors:  D Fiorella; P Lylyk; I Szikora; M E Kelly; F C Albuquerque; C G McDougall; P K Nelson
Journal:  J Neurointerv Surg       Date:  2009-06-16       Impact factor: 5.836

10.  Surgical treatment of giant intracranial aneurysms: current viewpoint.

Authors:  Giampaolo Cantore; Antonio Santoro; Giulio Guidetti; Catia P Delfinis; Claudio Colonnese; Emiliano Passacantilli
Journal:  Neurosurgery       Date:  2008-10       Impact factor: 4.654

View more
  7 in total

1.  Microsurgical strategies following failed endovascular treatment with the pipeline embolization device: case of a giant posterior cerebral artery aneurysm.

Authors:  Dale Ding; Robert M Starke; Kenneth C Liu
Journal:  J Cerebrovasc Endovasc Neurosurg       Date:  2014-03-31

2.  Balloon anchor technique for pipeline embolization device deployment across the neck of a giant intracranial aneurysm.

Authors:  Dale Ding; Robert M Starke; Avery J Evans; Mary E Jensen; Kenneth C Liu
Journal:  J Cerebrovasc Endovasc Neurosurg       Date:  2014-06-30

3.  Novel balloon application for rescue and realignment of a proximal end migrated pipeline flex embolization device into the aneurysmal sac: complication management.

Authors:  Mario Martínez-Galdámez; Joaquin Ortega-Quintanilla; Antonio Hermosín; Eduardo Crespo-Vallejo; Juan José Ailagas; Santiago Pérez
Journal:  BMJ Case Rep       Date:  2016-03-23

4.  Double-barrel Y-configuration Stenting for Flow Diversion of a Giant Recurrent Basilar Apex Aneurysm with the Pipeline Flex Embolization Device.

Authors:  Dale Ding; Robert M Starke; Ben McGuinness; Stefan Brew
Journal:  J Neurosci Rural Pract       Date:  2016-12

5.  Endovascular Management of Intracranial Aneurysms: Advances in Stenting Techniques and Technology.

Authors:  Dale Ding
Journal:  J Cerebrovasc Endovasc Neurosurg       Date:  2015-12-31

6.  Inadvertent Complication of a Pipeline Embolization Device for Treatment with Vertebral Artery Dissecting Aneurysm : Distal Tip Fracture of Delivery Wire.

Authors:  Jung Soo Park; Hyo Sung Kwak; Jong Myong Lee
Journal:  J Korean Neurosurg Soc       Date:  2016-09-08

7.  Endovascular Rescue Strategies for Nonopening of Pipeline Device: Report of Two Cases.

Authors:  Anshu Mahajan; Vinit Banga; Apratim Chatterjee; Gaurav Goel
Journal:  Asian J Neurosurg       Date:  2019-11-25
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.