Interface pressure in transtibial socket during ascent and descent on stairs and its effect on patient satisfaction.

Sadeeq Ali1, Noor Azuan Abu Osman, Arezoo Eshraghi, Hossein Gholizadeh, Nasrul Anwar Bin Abd Razak, Wan Abu Bakar Bin Wan Abas.   

Abstract

BACKGROUND: Transtibial amputees encounter stairs and steps during their daily activities. The excessive pressure between residual limb/socket may reduce the walking capability of transtibial prosthetic users during ascent and descent on stairs. The purposes of the research were to evaluate the interface pressure between Dermo (shuttle lock) and Seal-In X5 (prosthetic valve) interface systems during stair ascent and descent, and to determine their satisfaction effects on users.
METHODS: Ten amputees with unilateral transtibial amputation participated in the study. Interface pressure was recorded with F-socket transducer (9811E) during stair ascent and descent at self-selected speed. Each participant filled in a questionnaire about satisfaction and problems encountered with the use of the two interface systems.
FINDINGS: The resultant mean peak pressure (kPa) was significantly lower for the Dermo interface system compared to that of the Seal-In X5 interface system at the anterior, posterior and medial regions during stair ascent (63.14 vs. 80.14, 63.14 vs. 90.44, 49.21 vs. 66.04, respectively) and descent (67.11 vs. 80.41, 64.12 vs. 88.24, 47.33 vs. 65.11, respectively). Significant statistical difference existed between the two interface systems in terms of satisfaction and problems encountered (P<0.05).
INTERPRETATION: The Dermo interface system caused less pressure within the prosthetic socket compared to the Seal-In X5 interface system during stair negotiation. The qualitative survey also showed that the prosthesis users experienced fewer problems and increased satisfaction with the Dermo interface system.
© 2013. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Amputee; Pressure; Prosthetic interface systems; Satisfaction; Transtibial prosthesis

Mesh:

Year:  2013        PMID: 24161521     DOI: 10.1016/j.clinbiomech.2013.09.004

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  12 in total

1.  A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.

Authors:  Michael A McGeehan; Peter G Adamczyk; Kieran M Nichols; Michael E Hahn
Journal:  J Rehabil Assist Technol Eng       Date:  2022-07-15

2.  Stakeholder perspectives for possible residual limb monitoring system for persons with lower-limb amputation.

Authors:  Lilly Tran; Ryan Caldwell; Matthew Quigley; Stefania Fatone
Journal:  Disabil Rehabil       Date:  2018-09-05       Impact factor: 3.033

3.  Textile-based Pressure Sensors for Monitoring Prosthetic-Socket Interfaces.

Authors:  Jordan Tabor; Talha Agcayazi; Aaron Fleming; Brendan Thompson; Ashish Kapoor; Ming Liu; Michael Lee; He Helen Huang; Alper Bozkurt; Tushar Ghosh
Journal:  IEEE Sens J       Date:  2021-01-21       Impact factor: 3.301

4.  Detection of Prosthetic Knee Movement Phases via In-Socket Sensors: A Feasibility Study.

Authors:  Amr M El-Sayed; Nur Azah Hamzaid; Kenneth Y S Tan; Noor Azuan Abu Osman
Journal:  ScientificWorldJournal       Date:  2015-04-05

5.  Comparative study between Dermo, Pelite, and Seal-In X5 liners: effect on patient's satisfaction and perceived problems.

Authors:  Sadeeq Ali; Noor Azuan Abu Osman; Nooranida Arifin; Hossein Gholizadeh; Nasrul Anwar Abd Razak; Wan Abu Bakar Wan Abas
Journal:  ScientificWorldJournal       Date:  2014-08-11

6.  Piezoelectric bimorphs' characteristics as in-socket sensors for transfemoral amputees.

Authors:  Amr M El-Sayed; Nur Azah Hamzaid; Noor Azuan Abu Osman
Journal:  Sensors (Basel)       Date:  2014-12-10       Impact factor: 3.576

7.  Electroencephalogram-Based Brain-Computer Interface and Lower-Limb Prosthesis Control: A Case Study.

Authors:  Douglas P Murphy; Ou Bai; Ashraf S Gorgey; John Fox; William T Lovegreen; Brian W Burkhardt; Roozbeh Atri; Juan S Marquez; Qi Li; Ding-Yu Fei
Journal:  Front Neurol       Date:  2017-12-15       Impact factor: 4.003

8.  An Inductive Sensing System to Measure In-Socket Residual Limb Displacements for People Using Lower-Limb Prostheses.

Authors:  Katrina M Henrikson; Ethan J Weathersby; Brian G Larsen; John C Cagle; Jake B McLean; Joan E Sanders
Journal:  Sensors (Basel)       Date:  2018-11-09       Impact factor: 3.576

Review 9.  Techniques for Interface Stress Measurements within Prosthetic Sockets of Transtibial Amputees: A Review of the Past 50 Years of Research.

Authors:  Ebrahim A Al-Fakih; Noor Azuan Abu Osman; Faisal Rafiq Mahmad Adikan
Journal:  Sensors (Basel)       Date:  2016-07-20       Impact factor: 3.576

Review 10.  Prosthesis satisfaction in lower limb amputees: A systematic review of associated factors and questionnaires.

Authors:  Erwin C Baars; Ernst Schrier; Pieter U Dijkstra; Jan H B Geertzen
Journal:  Medicine (Baltimore)       Date:  2018-09       Impact factor: 1.889

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.