Diamond shaped optical fiber delivery system.

Haim Mnitentag1, Alexander Gersten, Shlomo Walfisch.   

Abstract

BACKGROUND: Optical fibers are generally used with a flat cut end. The adaptation of a coaxial exit of a refrigerated gas near the distal tip of the optical fiber cools the tissue where the laser beam is applied and reduces the injury of thermal effects on biological tissue.
MATERIALS AND METHODS: We have changed the flat end geometry of the fiber's end to a pyramidal end shape (named: Diamond tip). In this project we have evaluated the new shape of the fiber's end using laboratory tests and in real conditions. We have examined the emission pattern in the air and in water as well as its performance in surgery.
RESULTS: With the new shape better results were obtained: higher power density, more utilizations of the same fiber (after sterilization) resulting in economical savings. The diamond tip is characterized by greater mechanical resistance then conical tip delivery systems. The greater resistance of the Diamond tip prevents optical fiber breakage during surgery. The delivered system can be reshaped by a trained hospital physician and reused. Our experience shows that delivery systems intended by the producer to one use only can be converted, by reshaping their tips, to highly performing delivery systems, which can be reused many times. In addition less power was needed, with less tissue penetration. The thermal effect was minimized and higher mechanical resistance was achieved, which allowed reusing the fiber more than six times. The special structure of the faces of the optical fiber diamond tip eased its sliding.
CONCLUSION: The Diamond tip is superior to other fiber end tips by having minimal thermal effects, higher mechanical resistance and by being reused many times.

Entities:  

Keywords:  Optical fiber; diamond tip; higher mechanical resistance; pyramidal end shape; reduced thermal effects; reduced tissue penetration

Year:  2011        PMID: 24155510      PMCID: PMC3806080          DOI: 10.5978/islsm.20.29

Source DB:  PubMed          Journal:  Laser Ther        ISSN: 0898-5901


  5 in total

1.  Thermographic study of laser on arteries.

Authors:  J Mnitentag; E F Marques; M P Ribeiro; G A Braga; M R Navarro; A B Veratti; E Armelin; R Macruz; A D Jatene
Journal:  Lasers Surg Med       Date:  1987       Impact factor: 4.025

2.  Laser prostatectomy performed with a right angle firing neodymium:YAG laser fiber at 40 watts power setting.

Authors:  J N Kabalin
Journal:  J Urol       Date:  1993-07       Impact factor: 7.450

3.  Laser surgery in enclosed spaces: a review.

Authors:  R Macruz; M P Ribeiro; J M Brum; C A Pasqualucci; J Mnitentag; D G Bozinis; E Marques; A D Jatene; L V D'ecourt; E Armelin
Journal:  Lasers Surg Med       Date:  1985       Impact factor: 4.025

4.  The contact Nd:YAG laser system in the treatment of bladder cancer: a preliminary report in 48 patients.

Authors:  H Washida; H Watanabe; H Sakagami; S Sasaki; N Daikuzono
Journal:  Lasers Surg Med       Date:  1987       Impact factor: 4.025

5.  [Nd:YAG laser for anorectal surgery: initial experience in Israel].

Authors:  S Walfisch; N Ohana; E Charuzi
Journal:  Harefuah       Date:  1994-01-02
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.