Literature DB >> 24155151

Metal release and speciation of released chromium from a biomedical CoCrMo alloy into simulated physiologically relevant solutions.

Yolanda Hedberg1, Inger Odnevall Wallinder.   

Abstract

The objective of this study was to investigate the extent of released Co, Cr(III), Cr(VI), and Mo from a biomedical high-carbon CoCrMo alloy exposed in phosphate-buffered saline (PBS), without and with the addition of 10 µM H2 O2 (PBS + H2 O2 ), and 10 g L(-1) bovine serum albumin (PBS + BSA) for time periods up to 28 days. Comparative studies were made on AISI 316L for the longest time period. No Cr(VI) release was observed for any of the alloys in either PBS or PBS + H2 O2 at open-circuit potential (no applied potential). However, at applied potentials (0.7 V vs. Ag/AgCl), Cr was primarily released as Cr(VI). Co was preferentially released from the CoCrMo alloy at no applied potential. As a consequence, Cr was enriched in the utmost surface oxide reducing the extent of metal release over time. This passivation effect was accelerated in PBS + H2 O2 . As previously reported for 316L, BSA may also enhance metal release from CoCrMo. However, this was not possible to verify due to the precipitation of metal-protein complexes with reduced metal concentrations in solution as a consequence. This was particularly important for Co-BSA complexes after sufficient time and resulted in an underestimation of metals in solution.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  bovine serum albumin; chromium; hydrogen peroxide; metal release; speciation

Mesh:

Substances:

Year:  2013        PMID: 24155151     DOI: 10.1002/jbm.b.33048

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  6 in total

1.  Biocompatibility of Four Common Orthopedic Biomaterials Following a High-Salt Diet: An In Vivo Study.

Authors:  Mathieu Lecocq; Cécile Bernard; Marie Solenne Felix; Jean-Marc Linares; Julien Chaves-Jacob; Patrick Decherchi; Erick Dousset
Journal:  Int J Mol Sci       Date:  2017-07-11       Impact factor: 5.923

2.  Metals and Metalloids Release from Orthodontic Elastomeric and Stainless Steel Ligatures: In Vitro Risk Assessment of Human Exposure.

Authors:  Aneta Olszewska; Anetta Hanć; Danuta Barałkiewicz; Piotr Rzymski
Journal:  Biol Trace Elem Res       Date:  2019-11-04       Impact factor: 3.738

3.  Can gamma irradiation during radiotherapy influence the metal release process for biomedical CoCrMo and 316L alloys?

Authors:  Zheng Wei; Jonathan Edin; Anna Emelie Karlsson; Katarina Petrovic; Inna L Soroka; Inger Odnevall Wallinder; Yolanda Hedberg
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2018-02-09       Impact factor: 3.368

Review 4.  Mechanisms of Adverse Local Tissue Reactions to Hip Implants.

Authors:  Felipe Eltit; Qiong Wang; Rizhi Wang
Journal:  Front Bioeng Biotechnol       Date:  2019-07-30

5.  Effect of the coexistence of albumin and H2O2 on the corrosion of biomedical cobalt alloys in physiological saline.

Authors:  Weichen Xu; Binbin Zhang; Lihui Yang; Qiancheng Ni; Yantao Li; Fei Yu
Journal:  RSC Adv       Date:  2019-10-15       Impact factor: 4.036

6.  Titanium Implant Impairment and Surrounding Muscle Cell Death Following High-Salt Diet: An In Vivo Study.

Authors:  Mathieu Lecocq; Marie-Solenne Felix; Jean-Marc Linares; Julien Chaves-Jacob; Patrick Decherchi; Erick Dousset
Journal:  PLoS One       Date:  2016-01-13       Impact factor: 3.240

  6 in total