Literature DB >> 24149669

Battery life of portable home ventilators: effects of ventilator settings.

Line Falaize1, Karl Leroux2, Hélène Prigent1, Bruno Louis3, Sonia Khirani4, David Orlikowski1, Brigitte Fauroux5, Frédéric Lofaso6.   

Abstract

BACKGROUND: The battery life (BL) of portable home ventilator batteries is reported by manufacturers. The aim of this study was to evaluate the effects of ventilator mode, breathing frequency, PEEP, and leaks on the BL of 5 commercially available portable ventilators.
METHODS: The effects of the ventilator mode (volume controlled-continuous mandatory ventilation [VC-CMV] vs pressure support ventilation [PSV]), PEEP 5 cm H2O, breathing frequency (10, 15, and 20 breaths/min), and leaks during both volume-targeted ventilation and PSV on the BL of 5 ventilators (Elisée 150, Monnal T50, PB560, Vivo 50, and Trilogy 100) were evaluated. Each ventilator was ventilated with a test lung at a tidal volume of 700 ml and an inspiratory time of 1.2 s in the absence of leaks.
RESULTS: Switching from PSV to VC-CMV or the addition of PEEP did not significantly change ventilator BL. The increase in breathing frequency from 10 to 20 breaths/min decreased the BL by 18 ± 11% (P = .005). Leaks were associated with an increase in BL during the VC-CMV mode (18 ± 20%, P = .04) but a decrease in BL during the PSV mode (-13 ± 15%, P = .04).
CONCLUSIONS: The BL of home ventilators depends on the ventilator settings. BL is not affected by the ventilator mode (VC-CMV or PSV) or the addition of PEEP. BL decreases with an increase in breathing frequency and during leaks with a PSV mode, whereas leaks increase the duration of ventilator BL during VC-CMV.
Copyright © 2014 by Daedalus Enterprises.

Entities:  

Keywords:  batteries; home ventilators; leaks; portable ventilators; pressure support; volume-targeted mode

Mesh:

Year:  2013        PMID: 24149669     DOI: 10.4187/respcare.02711

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  2 in total

1.  Mouthpiece ventilation: Individualized patient care is the key to success.

Authors:  Adam Ogna; Frederic Lofaso
Journal:  Chron Respir Dis       Date:  2016-08-20       Impact factor: 2.444

2.  Bench-test comparison of 26 emergency and transport ventilators.

Authors:  Erwan L'Her; Annie Roy; Nicolas Marjanovic
Journal:  Crit Care       Date:  2014-10-15       Impact factor: 9.097

  2 in total