Sir, We read the article “Outcome of locking compression plates in humeral shaft nonunions” by Kumar et al.1 with keen interest. We would like to comment about a few discrepancies The authors have quoted a study by Ring et al.2 It is mentioned by the authors that “Ring et al. had reported average interval of 28 months (range 5-192 months),” but we did not find these findings in the original article by Ring et al.2 In the discussion the authors have mentioned that they inserted screws in all the holes of locking compression plate. They believed that reduced working length can minimize the failure of implant. However, literature reports mention that reduced working length can cause more strain on the implant and, hence, leads to plate failure.34
  4 in total

Review 1.  Guidelines for the clinical application of the LCP.

Authors:  Emanuel Gautier; Christoph Sommer
Journal:  Injury       Date:  2003-11       Impact factor: 2.586

Review 2.  Locking plates: tips and tricks.

Authors:  Wade R Smith; Bruce H Ziran; Jeff O Anglen; Philip F Stahel
Journal:  J Bone Joint Surg Am       Date:  2007-10       Impact factor: 5.284

3.  Locking compression plates for osteoporotic nonunions of the diaphyseal humerus.

Authors:  David Ring; Peter Kloen; John Kadzielski; David Helfet; Jesse B Jupiter
Journal:  Clin Orthop Relat Res       Date:  2004-08       Impact factor: 4.176

4.  Outcome of locking compression plates in humeral shaft nonunions.

Authors:  Malhar N Kumar; V Pratap Ravindranath; Mr Ravishankar
Journal:  Indian J Orthop       Date:  2013-03       Impact factor: 1.251

  4 in total
  1 in total

1.  Author's reply.

Authors:  Mn Kumar; V Pratap Ravindranath; Mr Ravishankar
Journal:  Indian J Orthop       Date:  2013-09       Impact factor: 1.251