Monitoring during hepatic radiofrequency ablation (RFA): comparison of real-time ultrasound elastography (RTE) and contrast-enhanced ultrasound (CEUS): first clinical results of 25 patients.

P Wiggermann1, K Brünn1, J Rennert1, M Loss2, H Wobser3, A G Schreyer1, C Stroszczynski1, E M Jung1.   

Abstract

PURPOSE: To evaluate the reliability of ultrasound elastography for delineating thermal ablation defects post-radiofrequency ablation (RFA) by comparing lesion dimensions determined by real-time elastography (RTE) with the findings of contrast-enhanced ultrasound (CEUS).
MATERIALS AND METHODS: A total of 21 malignant liver tumors were percutaneously ablated using RFA. Color-coded elastography and CEUS were performed by one experienced examiner, using a 1 - 5 MHz multi-frequency convex transducer (LOGIQ E9, GE). Lesions were examined using CEUS and real-time elastography (RTE) to assess ablation defects. Measurements of lesions (long axis, short axis, and area) representing the same image plane used for elastography were taken during CEUS examination and compared to the measurements obtained from the elastograms. All measurements were performed by two independent observers.
RESULTS: A statistically significant correlation in vivo between RTE and CEUS measurements with respect to the lesion's principal axis and area (r = 0.876 long axis, r = 0.842 short axis and r = 0.889 area) was found. Inter-rater reliability assessed with the concordance correlation coefficient was substantial for all measurements (ρc ≥ 0.96) Overall, elastography slightly underestimated the lesion size, as judged by the CEUS images.
CONCLUSION: These results support that RTE could potentially be used for the routine assessment of thermal ablation therapies. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24132649     DOI: 10.1055/s-0033-1355820

Source DB:  PubMed          Journal:  Ultraschall Med        ISSN: 0172-4614            Impact factor:   6.548


  6 in total

Review 1.  Ultrasound elastographic techniques in focal liver lesions.

Authors:  Clara Benedetta Conti; Federica Cavalcoli; Mirella Fraquelli; Dario Conte; Sara Massironi
Journal:  World J Gastroenterol       Date:  2016-03-07       Impact factor: 5.742

2.  Optimized Echo Decorrelation Imaging Feedback for Bulk Ultrasound Ablation Control.

Authors:  Mohamed A Abbass; Allison-Joy Garbo; Neeraja Mahalingam; Jakob K Killin; T Douglas Mast
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-06-14       Impact factor: 2.725

3.  Percutaneous irreversible electroporation for breast tissue and breast cancer: safety, feasibility, skin effects and radiologic-pathologic correlation in an animal study.

Authors:  Sheng Li; Fei Chen; Lujun Shen; Qi Zeng; Peihong Wu
Journal:  J Transl Med       Date:  2016-08-05       Impact factor: 5.531

4.  Thermographic real-time-monitoring of surgical radiofrequency and microwave ablation in a perfused porcine liver model.

Authors:  Florian Primavesi; Stefan Swierczynski; Eckhard Klieser; Tobias Kiesslich; Tarkan Jäger; Romana Urbas; Jörg Hutter; Daniel Neureiter; Dietmar Öfner; Stefan Stättner
Journal:  Oncol Lett       Date:  2017-12-18       Impact factor: 2.967

5.  In vivo ultrasound thermal ablation control using echo decorrelation imaging in rabbit liver and VX2 tumor.

Authors:  Mohamed A Abbass; Syed A Ahmad; Neeraja Mahalingam; K Sameer Krothapalli; Jack A Masterson; Marepalli B Rao; Peter G Barthe; T Douglas Mast
Journal:  PLoS One       Date:  2019-12-05       Impact factor: 3.240

6.  Hepatocellular adenoma: comparison between real-time contrast-enhanced ultrasound and dynamic computed tomography.

Authors:  Wei Wang; Jin-Ya Liu; Zheng Yang; Yue-Feng Wang; Shun-Li Shen; Feng-Lian Yi; Yang Huang; Er-Jiao Xu; Xiao-Yan Xie; Ming-De Lu; Zhu Wang; Li-Da Chen
Journal:  Springerplus       Date:  2016-06-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.