| Literature DB >> 24130840 |
Ping Pan1, Xiaoli Chen, Yu Li, Qingxue Zhang, Xiaomiao Zhao, Madafeitom M A Bodombossou-Djobo, Dongzi Yang.
Abstract
This study evaluated the consistency of manual and automated measurements of monodominant follicle diameter with different follicle size in infertile patients. Transvaginal two-dimensional (2D) ultrasound and SonoAVC (Sonography-based Automated Volume Calculation) were both performed in 226 infertile patients with monodominant follicle growth. 2D diameters were separately compared with SonoAVC-generated d(V) and m-d values in different follicle category, i.e. >10 to 14 mm, >14 to 18 mm, >18 to 22 mm and >22 mm. There was moderate degree of consistency between 2D diameter and SonoAVC-generated parameters regardless of follicle size. The mean differences were 0.82 mm between 2D diameter and SonoAVC-generated d(V) value, and 0.22 mm between 2D diameter and SonoAVC-generated m-d value, respectively. The discrepancy of manual and automated measurements tended to increase as follicle size increased. Our study suggested that compared with manual measurement, SonoAVC might underestimate follicle size. The absolute size of a follicle affected the consistency of two techniques.Entities:
Mesh:
Year: 2013 PMID: 24130840 PMCID: PMC3794997 DOI: 10.1371/journal.pone.0077095
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Correlation coefficients of 2D follicular diameter and SonoAVC-generated parameters in various follicle category.
| Follicle size (mm) | |||||
|---|---|---|---|---|---|
| Ultrasonic parameters | Total (n=312) | >10 to 14 (n=46) | >14 to 18 (n=107) | >18 to 22 (n=118) | >22 (n=41) |
| 2D diameter versus SonoAVC d(V) | 0.960 | 0.775 | 0.788 | 0.740 | 0.743 |
| 2D diameter versus SonoAVC m-d | 0.958 | 0.738 | 0.775 | 0.723 | 0.718 |
| 2D diameter versus SonoAVC V | 0.929 | 0.744 | 0.773 | 0.745 | 0.762 |
| SonoAVC d(V) versus SonoAVC m-d | 0.995 | 0.974 | 0.980 | 0.971 | 0.974 |
The ICCs of 2D measurement and SonoAVC in various follicle category.
| Follicle size (mm) | |||||
|---|---|---|---|---|---|
| Ultrasonic parameters | Total (n=312) | >10 to 14 (n=46) | >14 to 18 (n=107) | >18 to 22 (n=118) | >22 (n=41) |
| 2D diameter versus SonoAVC d(V) | |||||
| ICCs | 0.933 | 0.602 | 0.665 | 0.587 | 0.585 |
| 95% confidence intervals | 0.710-0.972 | 0.001-0.834 | 0.213-0.836 | 0.136-0.786 | 0.107-0.805 |
| 2D diameter versus SonoAVC m-d | |||||
| ICCs | 0.955 | 0.711 | 0.742 | 0.679 | 0.687 |
| 95% confidence intervals | 0.943-0.965 | 0.519-0.832 | 0.642-0.817 | 0.567-0.766 | 0.487-0.819 |
ICCs = intraclass correlation coefficients
Bland-Altman plot analysis of 2D measurement and SonoAVC in various follicle category.
| Follicle size (mm) | |||||
|---|---|---|---|---|---|
| Ultrasonic parameters | Total (n=312) | >10 to 14 (n=46) | >14 to 18 (n=107) | >18 to 22 (n=118) | >22 (n=41) |
| 2D diameter versus SonoAVC d(V) | |||||
| Mean difference | 0.82±0.95 | 0.68±0.62 | 0.77±0.92 | 0.86±1.00 | 1.02±1.19 |
| Upper limit of agreement | -1.08 | 1.92 | 2.61 | 2.86 | 3.40 |
| Lower limit of agreement | 2.72 | -0.56 | -1.07 | -1.14 | -1.36 |
| 2D diameter versus SonoAVC m-d | |||||
| Mean difference | 0.22±1.01 | 0.24±0.66 | 0.20±1.00 | 0.21±1.07 | 0.26±1.18 |
| Upper limit of agreement | -1.80 | 1.56 | 2.20 | 2.35 | 2.62 |
| Lower limit of agreement | 2.24 | -1.08 | -1.80 | -1.93 | -2.10 |