B W Jervis, M J Nichols, E M Allen, N R Hudson, T E Johnson.
Abstract
A quantitative assessment of both computerised correlation and analogue techniques for the removal of eye movement artefact from the electroencephalogram was undertaken. In both methods a fraction of the measured EOGs was subtracted from the measured EEG to leave the corrected EEG. In the correlation method the correction factors were computed from the cross-correlations of the EOGs and measured EEGs. In the analogue method the fraction subtracted was derived using a potential divider arrangement. The relative effectiveness of the two methods was determined by comparing the autocorrelation functions of the corrected EEGs at a lag of 2 sec. It was found that the computerised correlation technique was superior. A 3-channel EOG correlation correction procedure was found adequate in which two horizontal and one vertical EOGs were used. The analogue technique was very time consuming and possibly erroneous.Entities:
Mesh:
Year: 1985 PMID: 2412798 DOI: 10.1016/0013-4694(85)91037-5
Source DB: PubMed Journal: Electroencephalogr Clin Neurophysiol ISSN: 0013-4694