Literature DB >> 24125978

Dosimetric impact of Acuros XB dose calculation algorithm in prostate cancer treatment using RapidArc.

Suresh Rana1, Kevin Rogers, Terry Lee, Daniel Reed, Christopher Biggs.   

Abstract

PURPOSE: The purpose of this study is to assess the dosimetric impact of Acuros XB dose calculation algorithm (AXB), in comparisons with Anisotropic Analytical Algorithm (AAA) calculations in prostate cancer treatment using RapidArc.
MATERIALS AND METHODS: A computed tomography (CT) dataset of low-risk prostate cancer patients treated at Arizona Center for Cancer Care was selected and contoured for prostate, seminal vesicles, and organs at risk (OARs)(rectum, bladder, and femur heads). Plans were created for 6 MV photon beam using RapidArc technique in Eclipse treatment planning system. Dose calculations were performed with AAA and AXB for same number of monitor units and identical beam setup. Mean and maximum doses to planning target volume (PTV) and OARs were analyzed. Additionally, minimum dose to PTV and V100 was analyzed. Finally, point-dose difference between planar dose distributions of AAA and AXB plans was investigated.
RESULTS: The highest dose difference was up to 0.43% (range: 0.05-0.43%, P> 0.05) for PTV and 1.98% (range: 0.22-1.98%, P> 0.05) for OARs with AAA predicting higher dose than AXB. The V100 values of AAA plans (95 %) and AXB plans (range: 93.1-97.9 %) had an average difference of 0.89 ± 1.47% with no statistical significance (P = 0.25411). The point-dose difference analysis showed that AAA predicted higher dose than AXB at significantly higher percentage (in average 94.15) of total evaluated points.
CONCLUSION: The dosimetric results of this study suggest that the AXB can perform the dose computation comparable to AAA in RapidArc prostate cancer treatment plans that are generated by a partial single-arc technique.

Entities:  

Mesh:

Year:  2013        PMID: 24125978     DOI: 10.4103/0973-1482.119328

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  9 in total

1.  Dosimetric impact of Acuros XB on cervix radiotherapy using RapidArc technique: a dosimetric study.

Authors:  Lalit Kumar; Vimal Kishore; Manindra Bhushan; Pawan Kumar; Rahul Lal Chaudhary
Journal:  Rep Pract Oncol Radiother       Date:  2021-08-12

2.  Dosimetric Influence of Acuros XB Dose-to-Medium and Dose-to-Water Reporting Modes on Carcinoma Cervix Using Intensity-Modulated Radiation Therapy and Volumetric RapidArc Technique.

Authors:  Lalit Kumar; Manindra Bhushan; Vimal Kishore; Rahul Lal Chowdhary; Soumitra Barik; Anurag Sharma; Munish Gairola
Journal:  J Med Phys       Date:  2022-02-18

3.  Impact of acuros XB algorithm in deep-inspiration breath-hold (DIBH) respiratory techniques used for the treatment of left breast cancer.

Authors:  Lalit Kumar; Vimal Kishore; Manindra Bhushan; Abhinav Dewan; Girigesh Yadav; Kothanda Raman; Gourav Kumar; Irfan Ahmad; Kundan S Chufal; Munish Gairola
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-27

4.  Dosimetric effects of the acuros XB and anisotropic analytical algorithm on volumetric modulated arc therapy planning for prostate cancer using an endorectal balloon.

Authors:  Taeryool Koo; Jin-Beom Chung; Keun-Yong Eom; Jin-Yong Seok; In-Ah Kim; Jae-Sung Kim
Journal:  Radiat Oncol       Date:  2015-02-22       Impact factor: 3.481

5.  The quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note.

Authors:  Craig A Elith; Shane E Dempsey; Fred Cao; Afrooz Farshadi; Helen M Warren-Forward
Journal:  J Med Radiat Sci       Date:  2014-10-29

6.  Verification of Acuros XB dose algorithm using 3D printed low-density phantoms for clinical photon beams.

Authors:  Rodolfo Zavan; Philip McGeachy; Joseph Madamesila; Jose-Eduardo Villarreal-Barajas; Rao Khan
Journal:  J Appl Clin Med Phys       Date:  2018-03-25       Impact factor: 2.102

7.  Validation of the RapidArc Delivery System Using a Volumetric Phantom as Per Task Group Report 119 of the American Association of Physicists in Medicine.

Authors:  Lalit Kumar; Girigesh Yadav; Vimal Kishore; Manindra Bhushan; Munish Gairola; Deepak Tripathi
Journal:  J Med Phys       Date:  2019 Apr-Jun

Review 8.  Proton Therapy for Prostate Cancer: Challenges and Opportunities.

Authors:  Darren M C Poon; Stephen Wu; Leon Ho; Kin Yin Cheung; Ben Yu
Journal:  Cancers (Basel)       Date:  2022-02-13       Impact factor: 6.639

9.  A comparative analysis of Acuros XB and the analytical anisotropic algorithm for volumetric modulation arc therapy.

Authors:  Raju P Srivastava; K Basta; Werner De Gersem; Carlos De Wagter
Journal:  Rep Pract Oncol Radiother       Date:  2021-06-09