Correction of B0-distortions in echo-planar-imaging-based perfusion-weighted MRI.

Jonas Vardal1, Raimo A Salo, Christopher Larsson, Anders M Dale, Dominic Holland, Inge Rasmus Groote, Atle Bjørnerud.   

Abstract

PURPOSE: To evaluate and quantify a scheme for correcting susceptibility artifacts in spin-echo echo-planar-imaging-based dynamic susceptibility contrast (DSC) perfusion MRI of high-grade gliomas at 3 Tesla.
MATERIALS AND METHODS: Sixteen patients with a total of 78 scans were studied. DSC-MRI images were corrected using a displacement map generated from opposite phase-encoding polarity images. Two methods were used for quantification in the correction: (i) linear regression of pixel-by-pixel comparisons, performed both globally and relative to the anterior and posterior commissure plane (AC-PC plane), of T2-weighted images with both corrected and uncorrected raw DSC images; and (ii) counting significant (>2.0) normalized cerebral blood volume (nCBV) pixels from perfusion maps in the tumor region of interest.
RESULTS: Sixty-four of 78 datasets showed significant differences in the coefficient of correlation (r2) values. The difference between corrected and uncorrected r2 values was positive in all but one patient. Correction of B0- distortion significantly improved r2 in slices around the AC-PC plane. In 62% of the datasets, we observed an increased number of significant pixels in the corrected nCBV maps; 36% showed more significant pixels in uncorrected nCBV maps; 1% showed no difference.
CONCLUSION: Distortion correction of DSC-MRI may provide improved accuracy compared with uncorrected data, especially for tumors located below the corpus callosum and near the frontal sinuses.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  DSC-MRI; EPI; artifacts; correction; distortion; glioma

Mesh:

Substances:

Year:  2013        PMID: 24123663     DOI: 10.1002/jmri.24213

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  DR-BUDDI (Diffeomorphic Registration for Blip-Up blip-Down Diffusion Imaging) method for correcting echo planar imaging distortions.

Authors:  M Okan Irfanoglu; Pooja Modi; Amritha Nayak; Elizabeth B Hutchinson; Joelle Sarlls; Carlo Pierpaoli
Journal:  Neuroimage       Date:  2014-11-26       Impact factor: 6.556

2.  A novel approach to measure local cerebral haematocrit using MRI.

Authors:  Fernando Calamante; André Ahlgren; Matthias J P van Osch; Linda Knutsson
Journal:  J Cereb Blood Flow Metab       Date:  2015-09-30       Impact factor: 6.200

3.  Application of dual-source CT perfusion imaging and MRI for the diagnosis of primary liver cancer.

Authors:  Dongwen Zhang; Aixia Xu
Journal:  Oncol Lett       Date:  2017-05-15       Impact factor: 2.967

4.  Application of contrast-enhanced ultrasound and enhanced CT in diagnosis of liver cancer and evaluation of radiofrequency ablation.

Authors:  Shuhong Kong; Xuewang Yue; Sheng Kong; Yujie Ren
Journal:  Oncol Lett       Date:  2018-06-05       Impact factor: 2.967

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.