Eric M Schearer, Yu-Wei Liao, Eric J Perreault, Matthew C Tresch, William D Memberg, Robert F Kirsch, Kevin M Lynch.
Abstract
We present a method for controlling a neuroprosthesis for a paralyzed human arm using functional electrical stimulation (FES) and characterize the errors of the controller. The subject has surgically implanted electrodes for stimulating muscles in her shoulder and arm. Using input/output data, a model mapping muscle stimulations to isometric endpoint forces measured at the subject's hand was identified. We inverted the model of this redundant and coupled multiple-input multiple-output system by minimizing muscle activations and used this inverse for feedforward control. The magnitude of the total root mean square error over a grid in the volume of achievable isometric endpoint force targets was 11% of the total range of achievable forces. Major sources of error were random error due to trial-to-trial variability and model bias due to nonstationary system properties. Because the muscles working collectively are the actuators of the skeletal system, the quantification of errors in force control guides designs of motion controllers for multi-joint, multi-muscle FES systems that can achieve arbitrary goals.Entities:
Mesh:
Year: 2013 PMID: 24122573 PMCID: PMC4034352 DOI: 10.1109/TNSRE.2013.2282903
Source DB: PubMed Journal: IEEE Trans Neural Syst Rehabil Eng ISSN: 1534-4320 Impact factor: 3.802